Corrosion of the head-neck junction of the femoral component in total hip arthroplasty has been associated with symptomatic adverse local tissue reactions, trunion fracture and elevated serum metal ions. An analysis of risk factors and treatment strategies for corrosion at this interface is lacking in the literature. We therefore performed a systematic review of AAOS proceedings, MEDLINE and EMBASE databases, and included our own case series. A total of 24 articles representing 776 cases of head-neck corrosion met inclusion criteria. The combination of large femoral head sizes and small taper dimensions comprised the majority of published corrosion cases. Revision to ceramic head and ceramic/polyethylene liner was the most commonly utilised treatment. Coating precipitation, mixed alloy coupling and head-neck modulus mismatch collectively appear to contribute to the corrosive process.Keywords
This publication has 30 references indexed in Scilit:
- Corrosion at the Head-Neck Taper as a Cause for Adverse Local Tissue Reactions After Total Hip ArthroplastyJournal of Bone and Joint Surgery, 2012
- Metal-on-Metal Local Tissue Reaction Is Associated With Corrosion of the Head Taper JunctionThe Journal of Arthroplasty, 2012
- Severe corrosion after malpositioning of a metallic head over the Morse taper of a cementless hip arthroplasty. A case reportOrthopaedics & Traumatology: Surgery & Research, 2012
- Mixing and matching causing taper wearThe Journal of Bone and Joint Surgery. British volume, 2012
- Adverse reaction to metal release from a modular metal-on-polyethylene hip prosthesisThe Journal of Bone and Joint Surgery. British volume, 2011
- High failure rates with a large-diameter hybrid metal-on-metal total hip replacementThe Journal of Bone and Joint Surgery. British volume, 2011
- Total Hip Arthroplasty with Ceramic-on-Ceramic Bearing Failure from Third-Body WearOrthopedics, 2011
- Corrosion of a Hip Stem With a Modular Neck Taper JunctionThe Journal of Arthroplasty, 2009
- Metal Ion Levels in the Blood of Patients After Hip Resurfacing: A Comparison Between Twenty-eight and Thirty-six-Millimeter-Head Metal-on-Metal ProsthesesJournal of Bone and Joint Surgery, 2008
- Polyethylene wear in cases using femoral stems of similar geometry, but different metals, porous layer, and modularityThe Journal of Arthroplasty, 1998