Implantation Success, Electrical Performance, and Safety of an Active Fixation Stylet‐Driven Lead for LBBAP in Clinical Practice: A Multicenter Experience
Open Access
- 27 December 2025
- journal article
- Published by Wiley in Journal of Cardiovascular Electrophysiology
- Vol. 37 (3) , 475-483
Abstract
Background: Left bundle branch area pacing (LBBAP) has emerged as a promising physiological pacing strategy. However, its feasibility and safety using conventional stylet‐driven leads (SDL) such as INGEVITY+ remain under investigation. Objective: To evaluate the clinical performance, safety, and follow‐up outcomes of LBBAP using the INGEVITY+ lead in a real‐world electrophysiology setting. Methods: From 2021 to 2024, 207 consecutive patients underwent LBBAP implant attempts using the INGEVITY+ lead across 19 centers. Of these, 146 had bradycardia indications, and 61 had heart failure indications. A control group of 200 patients received standard right ventricular (RV) pacing with the same lead. Procedural outcomes, electrical parameters, complications, and mid‐term lead performance were assessed. Results: LBBAP was successfully achieved in 201 (97%) cases. Procedural and fluoroscopy times were slightly longer for LBBAP than for RV pacing. Acute complications included one helix fracture, one atrioventricular block, and one septal perforation. During a median follow‐up of 9 months, no lead fractures were reported; lead dislodgment occurred in three cases. Kaplan–Meier analysis showed no significant difference in time to first lead‐related complication between the LBBAP and RV pacing groups (hazard ratio: 4.69, 95%CI: 0.63–34.90, p = 0.139). Electrical performance remained stable, with 98% of retained leads maintaining capture thresholds ≤ 2 V and 92% with sensed amplitudes ≥ 5 mV. Conclusions: LBBAP using the INGEVITY + SDL is feasible and safe in clinical practice, with excellent implant success and stable mid‐term electrical performance. These findings support its use as a viable conduction system pacing strategy.Keywords
All Related Versions
This publication has 21 references indexed in Scilit:
- Left Ventricular Mechanical Insights Into Left Bundle Branch Pacing and Left Ventricular Septal PacingJACC: Clinical Electrophysiology, 2025
- Performance of an active fixation stylet-driven lead in left bundle branch area pacing: Results from INSIGHT-LBBAHeart Rhythm, 2025
- Real-world adoption of left bundle branch area pacing: Insights from the Conduction-System pacing Italian Network Group (C-SING)International Journal of Cardiology, 2024
- Left bundle branch area pacing using a stylet-driven, retractable-helix lead: Short-term results from a prospective multicenter IDE trial (the BIO-CONDUCT study)Heart Rhythm, 2024
- Inadvertent septal perforation during conduction system pacing device implant: a case reportEuropean Heart Journal - Case Reports, 2024
- Lead performance of stylet-driven leads in left bundle branch area pacing: Results from a large single-center cohort and insights from in vitro bench testingHeart Rhythm, 2024
- Acute performance of stylet driven leads for left bundle branch area pacing: A comparison with lumenless leadsHeart Rhythm, 2023
- 2023 HRS/APHRS/LAHRS guideline on cardiac physiologic pacing for the avoidance and mitigation of heart failureJournal of Arrhythmia, 2023
- EHRA clinical consensus statement on conduction system pacing implantation: executive summary. Endorsed by the Asia-Pacific Heart Rhythm Society (APHRS), Canadian Heart Rhythm Society (CHRS) and Latin-American Heart Rhythm Society (LAHRS)EP Europace, 2023
- Left bundle branch area pacing outcomes: the multicentre European MELOS studyEuropean Heart Journal, 2022