Systematic evaluation of surgical insertion of flexible neural probe arrays into deeper brain targets using length modulation methods
Top Cited Papers
Open Access
- 14 January 2026
- journal article
- Published by IOP Publishing in Journal of Neural Engineering
- Vol. 23 (1)
Abstract
Objective. Penetrating polymer-based microelectrode arrays (pMEAs) offer the potential for long term high-quality electrophysiological recordings of dynamic neural activity. Compared to rigid metal wire and silicon MEAs, improved device-tissue interface stability has been reported. However, accurate surgical placement of long, thin shanks in deeper brain regions is challenging as flexibility is achieved at the expense of axial stiffness. This study systematically evaluates then compares two pMEA placement strategies—dissolvable dip coating and molded brace, both with bare, exposed pMEA tips— to address the need for consistent, reliable, and accurate surgical targeting. These methods were selected based on the criteria of ease of fabrication, surgical feasibility, and mechanical performance. Approach. Sham (mechanical model with no electrodes) and fully functional pMEAs with shanks up to 5.5 mm long were fabricated and then modified using biodegradable polyethylene glycol (PEG) to support implantation. PEG was applied to shanks by motorized dip coating or a mechanical mold. Dissolution time and insertion in agarose gel brain models and rat cortex were evaluated followed by targeting of dip coated pMEAs to the rat hippocampus. Main results. Dip coating at high withdrawal speeds achieved uniform coating on shanks. Both strategies yielded similar critical buckling forces and insertion forces for single shank and arrayed pMEAs. Dip coated pMEAs were successfully placed in hippocampal regions without severe tissue damage as confirmed by histology and recordings obtained. Significance. Dip coating is a simpler method to prepare pMEAs for surgical targeting of deep brain regions compared to the bracing technique, as it does not require both a specialized mold and application process. This work provides a guide for researchers using single or multi-shank pMEAs to an accessible insertion strategy for implanting into deep brain regions in rodents and other small animal models.Keywords
Funding Information
- University of Southern California Annenberg Fellowship
- National Institutes of Health (U01 NS099703)
- National Institutes of Health (U01 NS126046)
- National Institutes of Health (U24 NS113647)
This publication has 80 references indexed in Scilit:
- Buckling‐Resistant and Trace‐Stacked (BRATS) Design Enables Aid‐Free Implantation of Flexible Multielectrode Array with Minimized Inflammatory Tissue ResponseAdvanced Functional Materials, 2025
- An Ultraflexible Electrode Array for Large‐Scale Chronic Recording in the Nonhuman Primate BrainAdvanced Science, 2023
- Quantitative Assessment of the Mechanical Properties of the Neural Interface2023
- Nanoelectronics for Minimally Invasive Cellular RecordingsAdvanced Functional Materials, 2019
- Transparent, Flexible, Penetrating Microelectrode Arrays with Capabilities of Single‐Unit ElectrophysiologyAdvanced Biosystems, 2019
- Ultra-miniature ultra-compliant neural probes with dissolvable delivery needles: design, fabrication and characterizationBiomedical Microdevices, 2016
- Sol-Gel Technologies for Glass Producers and Users2004
- Stress in thermally annealed parylene filmsJournal of Electronic Materials, 1992
- A method for pneumatically inserting an array of penetrating electrodes into cortical tissueAnnals of Biomedical Engineering, 1992
- Dragging of a Liquid by a Moving Plate1988