Poisson Statistical Model of Ultrasound Super-Resolution Imaging Acquisition Time
- 17 May 2019
- journal article
- research article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control
- Vol. 66 (7) , 1246-1254
Abstract
A number of acoustic super-resolution techniques have recently been developed to visualize microvascular structure and flow beyond the diffraction limit. A crucial aspect of all ultrasound (US) super-resolution (SR) methods using single microbubble localization is time-efficient detection of individual bubble signals. Due to the need for bubbles to circulate through the vasculature during acquisition, slow flows associated with the microcirculation limit the minimum acquisition time needed to obtain adequate spatial information. Here, a model is developed to investigate the combined effects of imaging parameters, bubble signal density, and vascular flow on SR image acquisition time. We find that the estimated minimum time needed for SR increases for slower blood velocities and greater resolution improvement. To improve SR from a resolution of λ/10 to λ/20 while imaging the microvasculature structure modeled here, the estimated minimum acquisition time increases by a factor of 14. The maximum useful imaging frame rate to provide new spatial information in each image is set by the bubble velocity at low blood flows (<;150 mm/s for a depth of 5 cm) and by the acoustic wave velocity at higher bubble velocities. Furthermore, the image acquisition procedure, transmit frequency, localization precision, and desired super-resolved image contrast together determine the optimal acquisition time achievable for fixed flow velocity. Exploring the effects of both system parameters and details of the target vasculature can allow a better choice of acquisition settings and provide improved understanding of the completeness of SR information.Keywords
Funding Information
- Engineering and Physical Sciences Research Council (EP/N015487/1, EP/N014855/1)
- Imperial College London (EP/L015226/1)
- Wellcome Trust (102431)
- Wellcome EPSRC Centre for Medical Engineering at King’s College London (WT 203148/Z/16/Z)
- National Institute for Health Research
This publication has 31 references indexed in Scilit:
- Ultrafast imaging in biomedical ultrasoundIEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 2014
- Acoustic super-resolution with ultrasound and microbubblesPhysics in Medicine & Biology, 2013
- A Statistical Model for the Quantification of Microbubbles in Destructive ImagingInvestigative Radiology, 2010
- The effect of photoswitching kinetics and labeling densities on super-resolution fluorescence imagingJournal of Biotechnology, 2010
- Microvascular and Macrovascular Complications of DiabetesClinical Diabetes, 2008
- Peripheral Arterial Disease in People With DiabetesDiabetes Care, 2003
- Pulse Inversion Imaging of Liver Blood FlowInvestigative Radiology, 2000
- Characteristics of SonoVue™Echocardiography, 1999
- Quantification of liver blood volume: comparison of ultra short ti inversion recovery echo planar imaging (ulstir‐epi), with dynamic 3d‐gradient recalled echo imagingMagnetic Resonance in Medicine, 1995
- Tumor Angiogenesis: A New Significant and Independent Prognostic Indicator in Early-Stage Breast CarcinomaJNCI: Journal of the National Cancer Institute, 1992