Dose calibration optimization and error propagation in polymer gel dosimetry
Authors:
- A JirasekAJA Jirasek
- M HiltsMHM Hilts
- 17 January 2014
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 59 (3) , 597-614
Abstract
This study reports on the relative precision, relative error, and dose differences observed when using a new full-image calibration technique in NIPAM-based x-ray CT polymer gel dosimetry. The effects of calibration parameters (e.g. gradient thresholding, dose bin size, calibration fit function, and spatial remeshing) on subsequent errors in calibrated gel images are reported. It is found that gradient thresholding, dose bin size, and fit function all play a primary role in affecting errors in calibrated images. Spatial remeshing induces minimal reductions or increases in errors in calibrated images. This study also reports on a full error propagation throughout the CT gel image pre-processing and calibration procedure thus giving, for the first time, a realistic view of the errors incurred in calibrated CT polymer gel dosimetry. While the work is based on CT polymer gel dosimetry, the formalism is valid for and easily extended to MRI or optical CT dosimetry protocols. Hence, the procedures developed within the work are generally applicable to calibration of polymer gel dosimeters.Keywords
This publication has 20 references indexed in Scilit:
- An x-ray CT polymer gel dosimetry prototype: I. Remnant artefact removalPhysics in Medicine & Biology, 2012
- A Novel Thin NIPAM Gel Cassette Dosimeter for Photon-Beam RadiotherapyPLOS ONE, 2012
- SU-E-T-93: A CT Polymer Gel Dosimetry System for End-To-End DosimetryMedical Physics, 2011
- Cosolvent-free polymer gel dosimeters with improved dose sensitivity and resolution for x-ray CT dose responsePhysics in Medicine & Biology, 2011
- Polymer gel dosimetry using x-ray computed tomography: investigation of the effect of reconstruction techniqueJournal of Physics: Conference Series, 2010
- Polymer gel dosimeters with enhanced sensitivity for use in x-ray CT polymer gel dosimetryPhysics in Medicine & Biology, 2010
- Polymer gel dosimetryPhysics in Medicine & Biology, 2010
- Monte Carlo simulation of RapidArc radiotherapy deliveryPhysics in Medicine & Biology, 2008
- Adaptive mean filtering for noise reduction in CT polymer gel dosimetryMedical Physics, 2007
- Technical considerations for implementation of x-ray CT polymer gel dosimetryPhysics in Medicine & Biology, 2005