Higher power achieves greater local impedance drop, shorter ablation time, and more transmural lesion formation in comparison to lower power in local impedance guided radiofrequency ablation of atrial fibrillation
- 2 August 2023
- journal article
- research article
- Published by Wiley in Journal of Cardiovascular Electrophysiology
- Vol. 34 (9) , 1869-1877
Abstract
Background: Since the local impedance (LI) of the ablation catheter reflects tissue characteristics, the efficacy of higher power (HP) compared to lower power (LP) in LI‐guided ablation may differ from other index‐guided ablations.Objective: This study aimed to assess the efficacy of HP ablation in LI‐guided ablation of atrial fibrillation (AF).Methods: A prospective observational study was conducted, enrolling patients undergoing de novo ablation for AF. Pulmonary vein isolation was performed using point‐by‐point ablation with a RHYTHMIA HDxTM Mapping System and an open‐irrigated ablation catheter with mini‐electrodes (IntellaNav MIFI OI). Ablation was stopped when the LI drop reached 30 ohms, three seconds after the LI plateaued, or when ablation time reached 30 s. To balance the baseline differences, a unique method was used in which the power was changed between HP (45 W to anterior wall/40 W to posterior wall) and LP (35 W/30 W) alternately for each adjacent point.Results: A total of 551 ablations in 10 patients were analyzed (HP, n = 276; LP, n = 275). The maximum LI drop was significantly larger (HP: 28.3 ± 5.4 vs. LP: 24.8 ± 6.3 ohm), and the time to minimum LI was significantly shorter (HP: 15.0 ± 6.3 vs. LP: 19.3 ± 6.6 s) in the HP setting. The unipolar electrogram analysis of three patients revealed that the electrogram indicating transmural lesion formation was observed more frequently in the HP setting.Conclusion: In LI‐guided ablation, the HP could achieve a larger LI drop and shorter time to minimum LI, which may result in more transmural lesion formation compared to a LP setting.Keywords
This publication has 20 references indexed in Scilit:
- High‐power short‐duration versus standard radiofrequency ablation: Insights on lesion metricsJournal of Cardiovascular Electrophysiology, 2018
- First clinical use of novel ablation catheter incorporating local impedance dataJournal of Cardiovascular Electrophysiology, 2018
- Impact of impedance decrease during radiofrequency current application for atrial fibrillation ablation on myocardial lesion and gap formationJournal of Arrhythmia, 2018
- Novel Measure of Local Impedance Predicts Catheter–Tissue Contact and Lesion FormationCirculation: Arrhythmia and Electrophysiology, 2018
- Atrial fibrillation ablation using very short duration 50 W ablations and contact force sensing cathetersJournal of Interventional Cardiac Electrophysiology, 2018
- Impedance-guided Radiofrequency Ablation: Using Impedance to Improve Ablation OutcomesJournal of Innovations in Cardiac Rhythm Management, 2017
- Relationship Between Catheter Contact Force and Radiofrequency Lesion Size and Incidence of Steam Pop in the Beating Canine HeartCirculation: Arrhythmia and Electrophysiology, 2014
- Local unipolar and bipolar electrogram criteria for evaluating the transmurality of atrial ablation lesions at different catheter orientations relative to the endocardial surfaceHeart Rhythm, 2010
- Indication of the radiofrequency induced lesion size by pre-ablation measurementsEP Europace, 2005
- Comparison of In Vivo Tissue Temperature Profile and Lesion Geometry for Radiofrequency Ablation With a Saline-Irrigated Electrode Versus Temperature Control in a Canine Thigh Muscle PreparationCirculation, 1995