Comparison of reliability and repeatability of corneal curvature assessment with six keratometers
- 1 November 2016
- journal article
- research article
- Published by Taylor & Francis in Clinical and Experimental Optometry
- Vol. 99 (6) , 583-589
Abstract
Keratometric methodology varies between instruments and the differences may have a clinical impact. We investigated the agreement and reproducibility of six keratometers. Keratometry was performed on 100 subjects at two separate sessions with IOLMaster 500, Pentacam, OPD scanner, Medmont E300, Javal-Schiøtz and TMS-5. A second observer assessed 30 subjects to determine inter-observer variability. A single individual was assessed on 10 separate sessions to determine intra-observer variability. Data were analysed using coefficient of variation (CV) and intra-class correlation coefficient (ICCC) for intra-observer variation. Inter-observer concordance was evaluated by the ICCC. Bland-Altman plots, Pearson's correlation coefficient and repeated measures analysis of variance were used to assess agreement of data produced by the instruments. OPD scanner and Javal-Schiøtz mean spherical equivalent (MSE) results were systematically different (p < 0.001) from other instruments (flatter and steeper, respectively). J0 /J45 were similar for all instruments (p < 0.05). Bland-Altman comparison plots indicated that Pentacam and IOLMaster demonstrated greatest level of agreement (ICC results MSE = 0.992, J0 = 0.934 and J45 = 0.890). Agreement (ICC) between observers for MSE ranged from 0.955 to 0.995 for all instruments; lower levels of agreement were found for J0 /J45 (0.289 to 0.901). IOLMaster showed greatest correlation and Medmont the lowest. All instruments showed high intra-observer repeatability of MSE (CV 0.1 to 0.3 per cent). The J0 /J45 readings showed greater variability (CV range 8.8 to 57.6 per cent). When considering MSE alone IOLMaster, Pentacam, OPD scan and Medmont may be considered interchangeable; however, assessment of astigmatism shows greater variability between instruments, sessions and observers.Keywords
This publication has 24 references indexed in Scilit:
- Comparison of corneal astigmatism measurements of 2 optical biometer models for toric intraocular lens selectionJournal of Cataract & Refractive Surgery, 2015
- Prediction of Residual Astigmatism After Cataract Surgery Using Swept Source Fourier Domain Optical Coherence TomographyCurrent Eye Research, 2014
- Comparison of Automated and Partial Coherence Keratometry and Resulting Choice of Toric IOLOptometry and Vision Science, 2013
- Comparison of corneal astigmatism and axis location in cataract patients measured by total corneal power, automated keratometry, and simulated keratometryJournal of Cataract & Refractive Surgery, 2012
- Comparability and repeatability of corneal astigmatism measurements using different measurement technologiesJournal of Cataract & Refractive Surgery, 2012
- Comparison of refractive outcomes using five devices for the assessment of preoperative corneal powerClinical & Experimental Ophthalmology, 2012
- Corneal topography with Scheimpflug imaging and videokeratography: Comparative study of normal eyesJournal of Cataract & Refractive Surgery, 2009
- The fourth New Zealand cataract and refractive surgery survey: 2007Clinical & Experimental Ophthalmology, 2008
- A new non-contact optical device for ocular biometryBritish Journal of Ophthalmology, 2002
- Toric intraocular lenses: Correcting astigmatism while controlling axis shiftJournal of Cataract & Refractive Surgery, 1994