Multiparameter Simulation for the Optimization of Ion Transmission Efficiency Through Mass Spectrometry Inlet Capillary
- 2 August 2026
- journal article
- research article
- Published by Wiley in Rapid Communications in Mass Spectrometry
- Vol. 40 (20) , e70155
Abstract
Rationale: According to the mass spectrometry (MS) design principles, the ions generated in an atmospheric pressure ion source must pass through a narrow‐bore capillary in order to maintain the needed system vacuum pressure. The ion transmission efficiency through the inlet capillary is one of the key factors directly influencing the analytical performance of MS. Its design and operating parameters, including inner diameter, length, temperature, and those related to the ion source operation, can alter ion diffusion losses, gas flow dynamics, and space‐charge effects both around and inside the inlet capillary, thereby influencing ion sampling efficiency and transmission efficiency to the mass analyzer. Methods: A comprehensive simulation of ion transmission efficiency through an MS inlet capillary has been performed in this study by adjusting the important design and operating parameters. Results: The simulation results show that the ion transmission efficiency through the inlet capillary is positively correlated with the inner diameter of the capillary and negatively correlated with the length of the capillary. Specifically, under the conditions of a capillary diameter of 0.8 mm and a temperature of 573.15 K, the ion transmission efficiency through a 10‐cm‐long capillary can reach 90%. The efficiency decreases with a reduction in the inner diameter or an increase in the length. When the temperature exceeds 573.15 K, the flow field inside the capillary transits from laminar to turbulent flow, leading to a sharp decline in ion transmission efficiency. For every 100‐K increase in temperature beyond this point, the ion transmission efficiency decreases by 10%–20%. Conclusions: The main results from this study have provided crucial design guidelines, emphasizing the use of larger diameters, moderate lengths, and strict temperature control below the turbulent transition threshold to maximize ion transmission and overall instrument sensitivity for MS instruments.Keywords
Funding Information
- National Key Research and Development Program of China (2023YFC3304203)
- K. C. Wong Magna Fund in Ningbo University (xkzwl1507)
This publication has 36 references indexed in Scilit:
- Ion transmission in an electrospray ionization‐mass spectrometry interface using an S‐lensJournal of Mass Spectrometry, 2023
- Investigation on choking behavior of gas flow in microductsMicrofluidics and Nanofluidics, 2018
- Gas Flow and Ion Transfer in Heated ESI Capillary InterfacesJournal of the American Society for Mass Spectrometry, 2018
- Gas Flow in the Capillary of the Atmosphere-to-Vacuum Interface of Mass SpectrometersJournal of the American Society for Mass Spectrometry, 2017
- Gas Flow Dynamics in Inlet Capillaries: Evidence for non Laminar ConditionsJournal of the American Society for Mass Spectrometry, 2016
- Numerical Modeling of Ion Transport in an ESI-MS SystemJournal of the American Society for Mass Spectrometry, 2014
- Study of the efficiency for ion transfer through bent capillariesJournal of Mass Spectrometry, 2012
- Capillary flow of low‐density polyethylenePolymer Engineering & Science, 2012
- Particles in Turbulent Flows2008
- Ion transport by viscous gas flow through capillariesJournal of the American Society for Mass Spectrometry, 1994