Comparison of specular microscopy and ultrasound pachymetry before and after cataract surgery
- 22 November 2016
- journal article
- research article
- Published by Springer Nature in Graefe's Archive for Clinical and Experimental Ophthalmology
- Vol. 255 (2) , 387-392
Abstract
To compare specular microscopy (Topcon SP-3000P, Topcon Corporation, Tokyo, Japan) and ultrasound (US) technology when evaluating central corneal thickness (CCT) prior to and after phacoemulsification. Corneal edema was assessed in phacoemulsification patients due to senile cataract by measuring CCT preoperatively and 1 day, 1 month, and 3 months postoperatively. Bland–Altman analysis was performed to assess interchangeability between pachymetry techniques for each visit. Repeated measures analysis of variance was performed to evaluate variation in CCT agreement depending on the degree of corneal edema. One hundred and eighteen patients aged 73.9 ± 10.1 years were recruited. Topcon SP-3000P provided significant (p < 0.0001) lower CCT values than US with and without induced corneal edema. Mean differences between CCT techniques for baseline, 1 day, 1 month and 3 months after cataract surgery were −28.9 ± 22.6, −30.5 ± 41.4, −32.3 ± 16.2 and −33.0 ± 16.9 μm, respectively. The systematic bias observed was not significantly different among the 4 visits (p = 0.59). The estimated limits of agreement (based on 1.96 standard deviation) were substantial, being 90.5, 165.8, 64.9 and 67.5 um at baseline, 1 day and 1 and 3 months. Topcon SP-3000P provides similar systematic bias in comparison with US technique for CCT measurements regardless of the degree of corneal edema after phacoemulsification, which should be reduced after applying a constant calibration adjustment of ∼30 μm. However, interchangeability between techniques may be still limited by the notable random measurement error.Keywords
Funding Information
- Instituto de Salud Carlos III (RD12/0034/0001)
This publication has 25 references indexed in Scilit:
- Dependability of Pachymetry Measurements after Myopic Advanced Surface Ablation Using Scanning-Slit Topography and Specular MicroscopyInvestigative Ophthalmology & Visual Science, 2013
- Evaluation of the Comparative Effect of Tetracaine on Central Corneal Thickness Measured by a Contact and Noncontact PachymeterJournal of Ocular Pharmacology and Therapeutics, 2013
- Repeatability and interobserver reproducibility of Artemis-2 high-frequency ultrasound in determination of human corneal thicknessClinical Ophthalmology, 2012
- Comparison of the precision of the Topcon SP-3000P specular microscope and an ultrasound pachymeterClinical Ophthalmology, 2011
- Statistical methods for conducting agreement (comparison of clinical tests) and precision (repeatability or reproducibility) studies in optometry and ophthalmologyOphthalmic and Physiological Optics, 2011
- Corneal oedema after cataract surgery: predisposing factors and corneal graft outcomeActa Ophthalmologica, 2009
- Comparison of central corneal thickness as measured by non‐contact specular microscopy and ultrasound pachymetry before and post LASIKClinical & Experimental Ophthalmology, 2007
- Extracellular Matrix and Na + ,K + -ATPase in Human Corneas Following Cataract SurgeryCornea, 2002
- Human Corneal Thickness and Its Impact on Intraocular Pressure MeasuresSurvey of Ophthalmology, 2000
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986