Dose reduction for chest CT: comparison of two iterative reconstruction techniques
- 1 June 2015
- journal article
- research article
- Published by SAGE Publications in Acta Radiologica
- Vol. 56 (6) , 688-695
Abstract
Background: Lowering radiation dose in computed tomography (CT) scan results in low quality noisy images. Iterative reconstruction techniques are used currently to lower image noise and improve the quality of images. Purpose: To evaluate lesion detection and diagnostic acceptability of chest CT images acquired at CTDIvol of 1.8 mGy and processed with two different iterative reconstruction techniques. Material and Methods: Twenty-two patients (mean age, 60 ± 14 years; men, 13; women, 9; body mass index, 27.4 ± 6.5 kg/m2) gave informed consent for acquisition of low dose (LD) series in addition to the standard dose (SD) chest CT on a 128 - multidetector CT (MDCT). LD images were reconstructed with SafeCT C4, L1, and L2 settings, and Safire S1, S2, and S3 settings. Three thoracic radiologists assessed LD image series (S1, S2, S3, C4, L1, and L2) for lesion detection and comparison of lesion margin, visibility of normal structures, and diagnostic confidence with SD chest CT. Inter-observer agreement (kappa) was calculated. Results: Average CTDIvol was 6.4 ± 2.7 mGy and 1.8 ± 0.2 mGy for SD and LD series, respectively. No additional lesion was found in SD as compared to LD images. Visibility of ground-glass opacities and lesion margins, as well as normal structures visibility were not affected on LD. CT image visibility of major fissure and pericardium was not optimal in some cases ( n = 5). Objective image noise in some low dose images processed with SafeCT and Safire was similar to SD images ( P value > 0.5). Conclusion: Routine LD chest CT reconstructed with iterative reconstruction technique can provide similar diagnostic information in terms of lesion detection, margin, and diagnostic confidence as compared to SD, regardless of the iterative reconstruction settings.Keywords
This publication has 21 references indexed in Scilit:
- Image quality assessment of ultra low-dose chest CT using sinogram-affirmed iterative reconstructionEuropean Radiology, 2014
- Sinogram-Affirmed Iterative Reconstruction of Low-Dose Chest CT: Effect on Image Quality and Radiation DoseAmerican Journal of Roentgenology, 2013
- CT dose and image quality in the last three scanner generationsWorld Journal of Radiology, 2013
- Pointers for Optimizing Radiation Dose in Chest CT ProtocolsJournal of the American College of Radiology, 2011
- Adaptive Statistical Iterative Reconstruction Technique for Radiation Dose Reduction in Chest CT: A Pilot StudyRadiology, 2011
- Chest computed tomography using iterative reconstruction vs filtered back projection (Part 2): image quality of low-dose CT examinations in 80 patientsEuropean Radiology, 2010
- A Prospective Evaluation of Dose Reduction and Image Quality in Chest CT Using Adaptive Statistical Iterative ReconstructionAmerican Journal of Roentgenology, 2010
- Scanning Beyond Anatomic Limits of the Thorax in Chest CT: Findings, Radiation Dose, and Automatic Tube Current ModulationAmerican Journal of Roentgenology, 2005
- Chest CT Performed with Z-Axis Modulation: Scanning Protocol and Radiation DoseRadiology, 2005
- Strategies for CT Radiation Dose OptimizationRadiology, 2004