Evidence of Reactivity in the Membrane for the Unstable Monochloramine during MIMS Analysis
Open Access
Abstract
Membrane Inlet Mass Spectrometry (MIMS) was used to analyze monochloramine solutions (NH2Cl) and ammonia solutions in a compact FTICR. Chemical ionization enables identification and quantification of the products present in the permeate. The responses of protonated monochloramine and ammonium increase linearly with the solution concentration. The enrichments were respectively 1.2 and 5.5. Pervaporation is dependent on pH and only the basic form of ammonia NH3 pervaporates through the membrane. Unexpectedly, the small ammonia molecule permeated very slowly. It could be due to interactions with water molecules inside the membrane that create clusters. Moreover, NH2Cl solutions, in addition to the NH3Cl+ signal, presented a strong NH4+ signal at m/z 18.034. Ammonia presence in the low-pressure zone before ionization is probable as NH4+ was detected with all the precursors used, particularly CF3+ and trimethylbenzene that presents a proton affinity higher than monochloramine. Ammonia may be formed inside the membrane due to the fact that NH2Cl is unstable and may react with the water present in the membrane. Those results highlight the need for caution when dealing with chloramines in MIMS and more generally with unstable molecules.Keywords
All Related Versions
Funding Information
- Minsitère de l'Économie et des Finances - France (RAPID 2013)
- Université Paris Sud (APP Attractivité 2011)
This publication has 33 references indexed in Scilit:
- Measuring 15N Abundance and Concentration of Aqueous Nitrate, Nitrite, and Ammonium by Membrane Inlet Quadrupole Mass SpectrometryAnalytical Chemistry, 2017
- A membrane introduction mass spectrometer utilizing ion‐molecule reactions for the on‐line speciation and quantitation of volatile organic moleculesRapid Communications in Mass Spectrometry, 2015
- Membrane introduction mass spectrometry (MIMS): a versatile tool for direct, real‐time chemical measurementsJournal of Mass Spectrometry, 2014
- Characterization of a membrane inlet interfaced with a compact chemical ionization FT-ICR for real-time and quantitative VOC analysis in waterInternational Journal of Mass Spectrometry, 2013
- Proton-Transfer Reaction Mass SpectrometryChemical Reviews, 2009
- Quantitative analysis of a complex mixture using proton transfer reaction in an FTICR mass spectrometerInternational Journal of Mass Spectrometry, 2008
- NEREUS: engineering concept for an underwater mass spectrometerTrAC Trends in Analytical Chemistry, 2002
- Underwater mass spectrometers for in situ chemical analysis of the hydrosphereJournal of the American Society for Mass Spectrometry, 2001
- Comparison of different methods for the determination of volatile organic compounds in water samplesTalanta, 1997
- Membrane interface for selective introduction of volatile compounds directly into the ionization chamber of a mass spectrometerAnalytical Chemistry, 1987