Abstract
We investigated protection strategies of organs at risk (OARs) in intensity modulated radiation therapy(IMRT). These strategies apply to delivery of IMRT to moving body anatomies that show relative displacement of OAR in close proximity to a tumor target. We formulated an efficient genetic algorithm which makes it possible to search for global minima in a complex landscape of multiple irradiation strategies delivering a given, predetermined intensity map to a target. The optimal strategy was investigated with respect to minimizing the dosedelivered to the OAR. The optimization procedure developed relies on variability of all parameters available for control of radiationdelivery in modern linear accelerators, including adaptation of leaf trajectories and simultaneous modification of beam dose rate during irradiation. We showed that the optimization algorithms lead to a significant reduction in the dosedelivered to OAR in cases where organs at risk move relative to a treatment target.

This publication has 11 references indexed in Scilit: