Electromagnetic Immunity of Pacemakers and Implantable Defibrillators to Wireless Power Transfer Systems for Automotive: A Provocative Study
- 12 December 2023
- journal article
- research article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Electromagnetic Compatibility
- Vol. 66 (1) , 97-107
Abstract
Wireless power transfer (WPT) in automotive is an emerging technology that utilizes coupled coils to transfer power through time-varying magnetic fields, enabling the charging of electric vehicles. Like any electromagnetic field source, WPT systems can potentially interfere with other electronic devices, impacting their proper functioning. This risk requires evaluation, especially when the affected device is a cardiac implantable electronic device (CIED), as malfunctioning can pose serious hazards to the patient's safety. Currently, WPT technology is in its early stages and has not been explicitly considered during the drafting of current CIED regulations. Consequently, predicting the behavior of a CIED when in proximity to a WPT system is not straightforward, and new experimental data are necessary. The objective of this article is to conduct in-vitro testing on a representative sample of pacemakers (PMs) and implantable cardioverter/defibrillators exposed to a magnetic field at the frequency of WTP systems, with field levels gradually increasing up to ICNIRP occupational reference levels. Compared to previous studies, the main novelty is to assess, on 11 commercial CIEDs, a safety margin for the exposure to the magnetic field dispersed by a WPT system, under worst-case exposure condition. Over a total of 51 conducted tests 24 interfering events were observed: only 1 occurred with a magnetic field at 50 μ T while all the others were observed with higher field levels, at 75 and 100 μ T. Thus, the results of this research demonstrate that the risk of interference is very low. Electromagnetic interference phenomena were observed only under worst-case conditions (e.g., maximum sensitivity of the device, maximum lead loop area, pulsed modulation, and unipolar configuration for PMs) and at exposure levels above the ICNIRP general population reference levels.Keywords
Funding Information
- Istituto Nazionale per l'Assicurazione Contro Gli Infortuni sul Lavoro
- Italian Workers’ Compensation Authority (36/2022)
This publication has 29 references indexed in Scilit:
- Review of Polynomial Chaos-Based Methods for Uncertainty Quantification in Modern Integrated CircuitsElectronics, 2018
- Human Exposure Assessment in Dynamic Inductive Power Transfer for Automotive ApplicationsIEEE Transactions on Magnetics, 2017
- Active implantable medical device EMI assessment for wireless power transfer operating in LF and HF bandsPhysics in Medicine & Biology, 2016
- Are patients with cardiac implants protected against electromagnetic interference in daily life and occupational environment?European Heart Journal, 2015
- Dosimetry of electromagnetic field exposure of an active armlet and its electromagnetic interference to the cardiac pacemakers using adult, child and infant modelsElectromagnetic Biology and Medicine, 2015
- Present Guidelines for Device ImplantationCirculation, 2014
- Interference between Active Implanted Medical Devices and Electromagnetic Field Emitting Devices is Rare but Real: Results of an Incidence Study in a Population of Physicians in FrancePacing and Clinical Electrophysiology, 2013
- Confirmation of quasi-static approximation in SAR evaluation for a wireless power transfer systemPhysics in Medicine & Biology, 2013
- GUIDELINES FOR LIMITING EXPOSURE TO TIME-VARYING ELECTRIC AND MAGNETIC FIELDS (1 Hz TO 100 kHz)Health Physics, 2010
- Global sensitivity analysis using polynomial chaos expansionsReliability Engineering & System Safety, 2008