Development, evaluation of critical method variables and stability assessment using a Box–Behnken design for the determination of organic impurities in a pharmaceutical dosage form of a centrally acting muscle relaxant drug chlorzoxazone
Open Access
- 4 September 2024
- journal article
- research article
- Published by Wiley in Biomedical Chromatography
- Vol. 38 (11) , e6001
Abstract
This study validates a stability‐indicating LC method for detecting organic impurities in the chlorzoxazone dosage form. Using a Waters X‐Select R HSS T3 analytical column, mobile phase of it was made by mixing of water, methanol, and glacial acetic acid in the ratio of 700:300:10 (v/v/v). The drug product and drug substance were subjected to the stress conditions such as acid, base, oxidation, heat, and photolysis as per the recommendations of the International Conference on Harmonization (Q2) methodology. The study revealed the susceptibility of 4‐chloro‐2‐aminophenol to alkaline environments, emphasizing peak homogeneity and stability. The method verification, per ICH guidelines and USP, established precision, specificity, linearity, accuracy, and robustness for quality control. The mean impurity recovery ranged from 95.5% to 105.2%, the correlation coefficient ( r ) was greater than 1.000, and the RSD values ( n = 6) ranged from 0.6% to 5.1% across the LOQ–150% ranges. Full‐factorial design tested final method conditions, evaluating multiple parameters concurrently. Graphical optimization within the design space defined strong method requirements, ensuring consistent and reliable outcomes. The study develops and validates chlorzoxazone stability‐indicating methods, employing advanced statistical approaches like design of experiments and factorial design, with resilient conditions established through graphical optimization of the design space.Keywords
This publication has 20 references indexed in Scilit:
- A quality‐by‐design evaluated liquid chromatography method development and validation for the separation and quantification of nitroxoline and its impuritiesJournal of Separation Science, 2023
- Stability indicating reversed‐phase‐high‐performance liquid chromatography method development and validation for pyridostigmine bromide and sodium benzoate in oral solutionBiomedical Chromatography, 2023
- Quality by design tool–evaluated stability‐indicating ultra‐performance liquid chromatography method for the determination of drugs (ritonavir and darunavir) used to treat the human immunodeficiency virus/acquired immunodeficiency syndromeBiomedical Chromatography, 2023
- An effective and stability‐indicating method development and optimization utilizing the Box–Behnken design for the simultaneous determination of acetaminophen, caffeine, and aspirin in tablet formulationBiomedical Chromatography, 2023
- QbD‐based stability‐indicating liquid chromatography (RP‐HPLC) method for the determination of flurbiprofen in cataplasmBiomedical Chromatography, 2023
- Degradation pathways and impurity profiling of the anticancer drug apalutamide by HPLC and LC–MS/MS and separation of impurities using Design of ExperimentsBiomedical Chromatography, 2022
- Determination of progesterone (steroid drug) in the semi‐solid dosage form (vaginal gel) using a stability‐indicating method by RP‐HPLC/PDA detectorBiomedical Chromatography, 2021
- Determination of Chlorzoxazone in Rat Plasma by LC-ESI-MS/MS and Its Application to a Pharmacokinetic StudyAnalytical Letters, 2010
- Development and Validation of RP-HPLC Method for Simultaneous Estimation of Three-Component Tablet Formulation Containing Acetaminophen, Chlorzoxazone, and AceclofenacAnalytical Letters, 2008
- A sensitive and rapid assay for 4-aminophenol in paracetamol drug and tablet formulation, by flow injection analysis with spectrophotometric detectionTalanta, 2002