Multi-channel bioimpedance spectroscopy based on orthogonal baseband shifting
- 1 June 2021
- journal article
- research article
- Published by IOP Publishing in Physiological Measurement
- Vol. 42 (6) , 064002
Abstract
Objective . Multi-channel bioimpedance spectroscopy (BIS) systems typically sample each channel’s impedance sequentially using multiplexers and a single analog-to-digital converter. These systems may lose their real-time capability with an increasing number of channels, especially for low excitation frequencies. We propose a new method, called orthogonal baseband shifting (OBS), for high-speed parallel BIS data acquisition at multiple excitation frequencies with low hardware and computational effort. Approach . Similar to orthogonal frequency-division multiplexing, used for digital data transmission, OBS systems use channel-specific orthogonal carrier frequencies to modulate the voltage response of the tissue. Given a suitable choice of carrier frequencies, the modulated signals of all channels sum up without loss of information and cross-talk. The fast Fourier transform (FFT) of the summed signal reveals a spectrum of non-overlapping, interleaved BIS data from which the corresponding BIS data of each channel can be calculated. Main results . In simulations, the system design requires a minimum signal-to-noise ratio of 30 dB to achieve amplitude errors below 1% and phase errors below 0.8°. The hardware realization, called ‘AixBIS’, has been evaluated for impedance measurements between 0.1 and 10 Ω with multi-frequency excitations between 45 and 180 kHz. The impedance values acquired had an averaged precision of 3.67 mΩ, which is only 0.65‰ in relation to the measured impedance. The phase had a mean precision of 0.46°. Moreover, in vitro measurements achieved 140 full spectrum acquisitions per second. The impedance change measured in a silicone heart phantom showed a high correlation of 0.83 with the ventricles volume change (flow). Significance . The proposed method enables very fast impedance acquisition of all channels. A complete measurement is performed in the time of a single FFT acquisition, which is equal to the resolution bandwidth of the FFT. In addition, portable and low-power multi-channel BIS devices profit from highly reduced hardware effort. The outstanding performance of OBS measurements with the AixBIS system have the potential for in vivo BIS measurements in real-time.Keywords
This publication has 20 references indexed in Scilit:
- An improved crest factor minimization algorithm to synthesize multisines with arbitrary spectrumPhysiological Measurement, 2015
- Fast Electrical Impedance Tomography Based on Code-Division-Multiplexing Using Orthogonal CodesIEEE Transactions on Instrumentation and Measurement, 2015
- Bioelectrical Impedance Methods for Noninvasive Health Monitoring: A ReviewJournal of Medical Engineering, 2014
- Crest factor optimization of the multisine waveform for bioimpedance spectroscopyPhysiological Measurement, 2014
- Code-Division-Multiplexed Electrical Impedance Tomography SpectroscopyIEEE Transactions on Biomedical Circuits and Systems, 2009
- Optimal electrode placement and frequency range selection for the detection of lung water using Bioimpedance Spectroscopy2018 40th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC), 2007
- Understanding the meaning of accuracy, trueness and precisionAccreditation and Quality Assurance, 2006
- Bioelectrical impedance analysis?part I: review of principles and methodsClinical Nutrition, 2004
- Continuous stroke volume and cardiac output from intra-ventricular dimensions obtained with impedance catheterCardiovascular Research, 1981
- A Mathematical Treatment of the Electric Conductivity and Capacity of Disperse Systems ii. The Capacity of a Suspension of Conducting Spheroids Surrounded by a Non-Conducting Membrane for a Current of Low FrequencyPhysical Review B, 1925