Optical imaging method to quantify spatial dose variation due to the electron return effect in an MR‐linac
- 10 December 2019
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 47 (3) , 1258-1267
Abstract
Purpose: Treatment planning systems (TPSs) for MR‐linacs must employ Monte Carlo‐based simulations of dose deposition to model the effects of the primary magnetic field on dose. However, the accuracy of these simulations, especially for areas of tissue‐air interfaces where the electron return effect (ERE) is expected, is difficult to validate due to physical constraints and magnetic field compatibility of available detectors. This study employs a novel dosimetric method based on remotely captured, real‐time optical Cherenkov and scintillation imaging to visualize and quantify the ERE.Methods: An intensified CMOS camera was used to image two phantoms with designed ERE cavities. Phantom A was a 40 cm × 10 cm × 10 cm clear acrylic block drilled with five holes of increasing diameters (0.5, 1, 2, 3, 4 cm). Phantom B was a clear acrylic block (25 cm × 20 cm × 5 cm) with three cavities of increasing diameter (3, 2, 1 cm) split into two halves in the transverse plane to accommodate radiochromic film. Both phantoms were imaged while being irradiated by 6 MV flattening filter free (FFF) beams within a MRIdian Viewray (Viewray, Cleveland, OH) MR‐linac (0.34 T primary field). Phantom A was imaged while being irradiated by 6 MV FFF beams on a conventional linac (TrueBeam, Varian Medical Systems, San Jose, CA) to serve as a control. Images were post processed in Matlab (Mathworks Inc., Natick, MA) and compared to TPS dose volumes.Results: Control imaging of Phantom A without the presence of a magnetic field supports the validity of the optical image data to a depth of 6 cm. In the presence of the magnetic field, the optical data shows deviations from the commissioned TPS dose in both intensity and localization. The largest air cavity examined (3 cm) indicated the largest dose differences, which were above 20% at some locations. Experiments with Phantom B illustrated similar agreement between optical and film dosimetry comparisons with TPS data in areas not affected by ERE.Conclusion: There are some appreciable differences in dose intensity and spatial dose distribution observed between the novel experimental data set and the dose models produced by the current clinically implemented MR‐IGRT TPS.Keywords
Funding Information
- National Institutes of Health (R01 EB023909, F31 CA192473, P30 CA 23108)
This publication has 44 references indexed in Scilit:
- Towards reference dosimetry for the MR-linac: magnetic field correction of the ionization chamber readingPhysics in Medicine & Biology, 2013
- Dosimetric evaluation of Acuros XB dose calculation algorithm with measurements in predicting doses beyond different air gap thickness for smaller and larger field sizesJournal of Medical Physics, 2013
- Quantitative imaging of electron density and effective atomic number using phase contrast CTPhysics in Medicine & Biology, 2010
- Polymer gel dosimetryPhysics in Medicine & Biology, 2010
- Dosimetry for the MRI accelerator: the impact of a magnetic field on the response of a Farmer NE2571 ionization chamberPhysics in Medicine & Biology, 2009
- Magnetic-field-induced dose effects in MR-guided radiotherapy systems: dependence on the magnetic field strengthPhysics in Medicine & Biology, 2008
- Experimental verification of magnetic field dose effects for the MRI-acceleratorPhysics in Medicine & Biology, 2007
- Integrating a MRI scanner with a 6 MV radiotherapy accelerator: impact of the surface orientation on the entrance and exit dose due to the transverse magnetic fieldPhysics in Medicine & Biology, 2007
- Integrating a MRI scanner with a 6 MV radiotherapy accelerator: dose increase at tissue–air interfaces in a lateral magnetic field due to returning electronsPhysics in Medicine & Biology, 2005
- Scintillating fiber dosimeter for radiation therapy acceleratorIEEE Transactions on Nuclear Science, 2002