Phase‐contrast MRI with hybrid one and two‐sided flow‐encoding and velocity spectrum separation
- 9 August 2016
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 78 (1) , 182-192
Abstract
Purpose: To develop and evaluate a phase‐contrast MRI (PC‐MRI) technique with hybrid one and two‐sided flow‐encoding and velocity spectrum separation (HOTSPA) for accelerated blood flow and velocity measurement.Methods: In the HOTSPA technique, the two‐sided flow encoding (FE) is used for two FE directions and one‐sided is used for the remaining FE direction. Such a temporal modulation of the FE strategy allows for separations of the Fourier velocity spectrum into components for the flow‐compensated and the three‐directional velocity waveforms, accelerating PC‐MRI by encoding three‐directional velocities using only two repetition times (TRs) instead of four TRs as in standard PC‐MRI. The HOTSPA was evaluated and compared with standard PC‐MRI in the common carotid arteries of six healthy volunteers.Results: Total volumetric flow and peak velocity measurements based on HOTSPA and the conventional PC‐MRI were in good agreement with a bias of −0.005 mL (−0.1% relative bias error) for total volumetric flow and 1.21 cm/s (1.1% relative bias error) for peak velocity, although the total acquisition time was 50% of the conventional PC‐MRI.Conclusion: The proposed HOTSPA technique achieved nearly two‐fold acceleration of PC‐MRI while maintaining accuracy for total volumetric flow and peak velocity quantification by separating the paired acquisitions in the Fourier velocity spectrum domain. Magn Reson Med 78:182–192, 2017. © 2016 International Society for Magnetic Resonance in MedicineKeywords
Funding Information
- National Heart, Lung, and Blood Institute (R01HL127153)
This publication has 26 references indexed in Scilit:
- Accelerated aortic flow assessment with compressed sensing with and without use of the sparsity of the complex difference imageMagnetic Resonance in Medicine, 2012
- Accelerated phase‐contrast cine MRI using k‐t SPARSE‐SENSEMagnetic Resonance in Medicine, 2011
- Highly k‐t‐space–accelerated phase‐contrast MRIMagnetic Resonance in Medicine, 2008
- Sparse MRI: The application of compressed sensing for rapid MR imagingMagnetic Resonance in Medicine, 2007
- Investigating myocardial motion by MRI using tissue phase mappingEuropean Journal of Cardio-Thoracic Surgery, 2006
- Controlled aliasing in parallel imaging results in higher acceleration (CAIPIRINHA) for multi‐slice imagingMagnetic Resonance in Medicine, 2005
- Accelerated dynamic Fourier velocity encoding by exploiting velocity-spatio-temporal correlationsMagnetic Resonance Materials in Physics, Biology and Medicine, 2004
- Generalized autocalibrating partially parallel acquisitions (GRAPPA)Magnetic Resonance in Medicine, 2002
- SENSE: Sensitivity encoding for fast MRIMagnetic Resonance in Medicine, 1999
- Analysis of encoding efficiency in MR imaging of velocity magnitude and directionMagnetic Resonance in Medicine, 1992