Can an increase in abutment screw length enable biomechanical advantages for implants with increased crown height space?
Open Access
- 15 May 2025
- journal article
- Published by Wiley in Journal of Prosthodontics
- Vol. 35 (4) , 544-552
Abstract
Purpose: The purpose of this finite element analysis (FEA) study was to investigate how screw length and crown height affect the stresses generated in the implant‐abutment complex and alveolar bone for anterior single implants and whether the use of longer abutment screw would compensate for increased stresses originated from increased crown height space (CHS). Materials and Methods: A maxillary anterior segment was modeled, and one implant (Straumann bone level implants, 4.1 mm × 12 mm) was placed at the central incisor site. Three different crown heights (11, 13, and 15 mm) with two different abutment screw lengths (regular [R] 7.9 and long [L] 9.9 mm) were modeled, and coded as M 11‐R, M 13‐R, M 15‐R, M 11‐L, M 13‐L, and M 15‐L depending on the height and length, respectively. Horizontal (25.5 N) and 30‐degree oblique forces (178 N) were applied to the crown. The stress magnitude and distribution were evaluated. Results: M 15‐L showed the highest von Mises stress value (869.72 MPa) at the abutment under oblique loads. For both screw lengths, von Mises stress values in the abutment, abutment screw, and inside and outside of the implant walls around the implant platform increased with increased crown height. With the increase in length of the abutment screw, the von Mises stresses decreased in the inside and outside walls of the implant, while the stresses in the abutment increased. Conclusions: Abutment screw length and crown height affected the amount of stress generated on implant components. Increasing the screw length may reduce the stresses on the implant and bone without increasing the stresses in the screw, potentially mitigating the risk of implant platform fractures and bone resorption, both of which are critical failures. Screw related issues may decrease as the area stresses effect has a tendency to reduce when longer screw is used in increased CHS situation (tall crowns).Keywords
This publication has 31 references indexed in Scilit:
- Effect of crown height on the screw joint stability of zirconia screw-retained crownsThe Journal of Prosthetic Dentistry, 2022
- Does the crown-implant ratio affect the survival and complications of implant-supported prostheses? A systematic reviewThe Journal of Prosthetic Dentistry, 2022
- The influence of crown-to-implant ratio in single crowns on clinical outcomes: A systematic review and meta-analysisThe Journal of Prosthetic Dentistry, 2021
- Impact of peri-implant bone resorption, prosthetic materials, and crown to implant ratio on the stress distribution of short implants: a finite element analysisMedical & Biological Engineering & Computing, 2021
- Biomechanical effect of implant design on four implants supporting mandibular full-arch fixed dentures: In vitro test and finite element analysisJournal of the Formosan Medical Association, 2020
- Effect of different restorative crown and customized abutment materials on stress distribution in single implants and peripheral bone: A three-dimensional finite element analysis studyThe Journal of Prosthetic Dentistry, 2018
- Analysis of the biomechanical behavior of short implants: The photo-elasticity methodMaterials Science and Engineering: C, 2015
- Effect of increased crown height on stress distribution in short dental implant components and their surrounding bone: A finite element analysisThe Journal of Prosthetic Dentistry, 2015
- Three-dimensional finite element analysis of stress distribution in retention screws of different crown–implant ratiosComputer Methods in Biomechanics and Biomedical Engineering, 2013
- Long‐Term Outcomes of Dental Implants with a Titanium Plasma‐Sprayed Surface: A 20‐Year Prospective Case Series Study in Partially Edentulous PatientsClinical Implant Dentistry and Related Research, 2013