Design and characterization of the time-resolved opacity spectrometer (OpSpecTR) for the NIF iron opacity campaign
- 1 August 2024
- journal article
- research article
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 95 (8)
Abstract
A new time-resolved opacity spectrometer (OpSpecTR) is currently under development for the National Ignition Facility (NIF) opacity campaign. The spectrometer utilizes Icarus version 2 (IV2) hybridized complementary metal–oxide–semiconductor sensors to collect gated data at the time of the opacity transmission signal, unlocking the ability to collect higher-temperature measurements on NIF. Experimental conditions to achieve higher temperatures are feasible; however, backgrounds will dominate the data collected by the current time-integrating opacity spectrometer. The shortest available OpSpecTR integration time of ∼2 ns is predicted to reduce self-emission and other late-time backgrounds by up to 80%. Initially, three Icarus sensors will be used to collect data in the self-emission, backlighter, and absorption regions of the transmission spectrum, with plans to upgrade to five Daedalus sensors in future implementations with integration times of ∼1.3 ns. We present the details of the diagnostic design along with recent characterization results of the IV2 sensors.Keywords
Funding Information
- U.S. Department of Energy (DE-AC52-07NA27344, 89233218CNA000001, DE-NA0003624)
This publication has 30 references indexed in Scilit:
- Progress toward NIF opacity measurementsHigh Energy Density Physics, 2020
- Superlattice-enhanced silicon soft X-ray and charged particle detectors with nanosecond time responseNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2019
- Replicating the Z iron opacity experiments on the NIFHigh Energy Density Physics, 2017
- Conceptual design of initial opacity experiments on the national ignition facilityJournal of Plasma Physics, 2017
- Design of the opacity spectrometer for opacity measurements at the National Ignition FacilityReview of Scientific Instruments, 2016
- A higher-than-predicted measurement of iron opacity at solar interior temperaturesNature, 2014
- High-energy x-ray backlighter spectrum measurements using calibrated image platesReview of Scientific Instruments, 2011
- Experimental investigation of opacity models for stellar interior, inertial fusion, and high energy density plasmasPhysics of Plasmas, 2009
- Helioseismology and solar abundancesPhysics Reports, 2008
- Quantitative measurement of mid-z opacitiesJournal of Quantitative Spectroscopy and Radiative Transfer, 1995