MRI Robot for Prostate Focal Laser Ablation: An Ex Vivo Study in Human Prostate
Authors:
- Reza SeifabadiRSReza Seifabadi
- Ming LiMLMing Li
- Sheng XuSXSheng Xu
- Yue ChenYue Chen
- Alex SquiresASAlex Squires
- Ayele H. NegussieANAyele H. Negussie
- Ivane BakhutashviliIBIvane Bakhutashvili
- Peter ChoykePCPeter Choyke
- Ismail B. TurkbeyITIsmail B. Turkbey
- Zion Tsz Ho TseZTZion Tsz Ho Tse
- Bradford J. WoodBWBradford J. Wood
Open Access
- 29 November 2018
- journal article
- research article
- Published by MDPI AG in Journal of Imaging
- Vol. 4 (12) , 140
Abstract
Purpose: A novel grid-template-mimicking MR-compatible robot was developed for in-gantry MRI-guided focal laser ablation of prostate cancer. Method: A substantially compact robot was designed and prototyped to meet in-gantry lithotomy ergonomics and allow for accommodation in the perineum. The controller software was reconfigured and integrated with the custom-designed navigation and multi-focal ablation software. Three experiments were conducted: (1) free space accuracy test; (2) phantom study under computed tomography (CT) guidance for image-guided accuracy test and overall workflow; and (3) magnetic resonance imaging (MRI)-guided focal laser ablation of an ex vivo prostate. The free space accuracy study included five targets that were selected across the workspace. The robot was then commanded five times to each target. The phantom study used a gel phantom made with color changing thermos-chromic ink, and four spherical metal fiducials were deployed with the robot. Then, laser ablation was applied, and the phantom was sliced for gross observation. For an MR-guided ex vivo test, a prostate from a donor who died of prostate cancer was obtained and multi-focally ablated using the system within the MRI gantry. The tissue was sliced after ablation for validation. Results: free-space accuracy was 0.38 ± 0.27 mm. The overall system targeting accuracy under CT guidance (including robot, registration, and insertion error) was 2.17 ± 0.47 mm. The planned ablation zone was successfully covered in both acrylamide gel phantom and in human prostate tissue. Conclusions: The new robot can accurately facilitate fiber targeting for MR-guided focal laser ablation of targetable prostate cancer.Keywords
This publication has 15 references indexed in Scilit:
- MRI Robots for Needle-Based Interventions: Systems and TechnologyAnnals of Biomedical Engineering, 2018
- Cancer statistics, 2018CA: A Cancer Journal for Clinicians, 2018
- Robotic System for MRI-Guided Focal Laser Ablation in the ProstateIEEE/ASME Transactions on Mechatronics, 2016
- Toward teleoperated needle steering under continuous MRI guidance for prostate percutaneous interventionsThe International Journal of Medical Robotics and Computer Assisted Surgery, 2015
- MRI-Safe Robot for Endorectal Prostate BiopsyIEEE/ASME Transactions on Mechatronics, 2014
- Treatment planning for prostate focal laser ablation in the face of needle placement uncertaintyMedical Physics, 2013
- Development and Preliminary Evaluation of a Motorized Needle Guide Template for MRI-Guided Targeted Prostate BiopsyIEEE Transactions on Biomedical Engineering, 2013
- An MRI-Compatible Robotic System With Hybrid Tracking for MRI-Guided Prostate InterventionIEEE Transactions on Biomedical Engineering, 2011
- MRI-Compatible Pneumatic Robot for Transperineal Prostate Needle PlacementIEEE/ASME Transactions on Mechatronics, 2008
- Robotic System for Closed-Bore MRI-Guided Prostatic InterventionsIEEE/ASME Transactions on Mechatronics, 2008