Omni-QALAS: Optimized Multiparametric Imaging for Simultaneous T1, T2 and Myelin Water Mapping

Authors:
  • UGUnay Dorken Gallastegi
    Unay Dorken Gallastegi
  • SFShohei Fujita
    Shohei Fujita
  • YCYuting Chen
    Yuting Chen
  • PXPengcheng Xu
    Pengcheng Xu
  • YCYangsean Choi
    Yangsean Choi
  • BGBorjan Gagoski
    Borjan Gagoski
  • HYHuihui Ye
    Huihui Ye
  • HLHuafeng Liu
    Huafeng Liu
  • BBBerkin Bilgic
    Berkin Bilgic

Abstract
Purpose: To improve the accuracy of multiparametric estimation, including myelin water fraction (MWF) quantification, and reduce scan time in 3D-QALAS by optimizing sequence parameters, using a self-supervised multilayer perceptron network. Methods: We jointly optimize flip angles, T2 preparation durations, and sequence gaps for T1 recovery using a self-supervised MLP trained to minimize a Cramer-Rao bound-based loss function, with explicit constraints on total scan time. The optimization targets white matter, gray matter, and myelin water tissues, and its performance was validated through simulation, phantom, and in vivo experiments. Results: Building on our previously proposed MWF-QALAS method for simultaneous MWF, T1, and T2 mapping, the optimized sequence reduces the number of readouts from six to five and achieves a scan time nearly one minute shorter, while also yielding higher T1 and T2 accuracy and improved MWF maps. This sequence enables simultaneous multiparametric quantification, including MWF, at 1 mm isotropic resolution within 3 minutes and 30 seconds. Conclusion: This study demonstrated that optimizing sequence parameters using a self-supervised MLP network improved T1, T2 and MWF estimation accuracy, while reducing scan time.

This publication has 41 references indexed in Scilit: