The Reliability between Placido Topography and Scheimpflug Topography in Normal Corneas with Astigmatism less Than 0.50 Dc
Authors:
Open Access
- 3 April 2020
- journal article
- research article
- Published by Sciencedomain International in Ophthalmology Research: An International Journal
- p. 27-35
Abstract
Purpose: This Clinical research was targeted to experimentally investigate the reliability between Placido topography and Scheimpflug topography in normal corneas with astigmatism ≤ ±0.50 Dc by comparing the refractive powers at the 4 main meridians (180⁰ - 90⁰) in order to study any differences between the two topographers and their reliability. Comparing the reliability measurements of Placido disk topography (CSO- Modi 02 Corneal Topographer with Phoenix Full & Pupil Module Software) with those of a high-resolution rotating Scheimpflug camera (Pentacam-Oculus,) and assess the agreement between these two topographers in measuring corneal power in normal eyes without astigmatism (≤ 0.50 Dc). Methods and Materials: Measurements were taken from 56 eyes corresponding to a proportion of patients, of which 26 were men with a mean age of ± 30 years and 30 were women with a mean age of ± 27 years. Refractive power maps were taken from the anterior cornea, having results for the steep meridian value (Ks), the flat meridian value (Kf), mean keratometric value (Km), and cornea astigmatism (Ks-Kf). The same examiner took all the data from measuring these 56 eyes. The results for each refractive map from the two topographers was taken with values for meridians (180⁰- 90⁰). Then the there was a statistical analysis comparing the values of (CSO) and (Pentacam) for those 56 eyes measured. Results: The statistical analysis for all meridians based on the results and as far as normal corneas with astigmatism ≤ ±0.50 Dc are concerned, the Placido topography (CSO) and Scheimpflug imaging technique (Pentacam), can be considered as equivalent in terms of diagnostic credibility and accuracy for primary health care, fitting contact lenses, etc. as the correlation coefficient r did not show any important differences for both instruments. For the 90⁰ meridian the correlation coefficient was 0.79, Significance level P=0.0001 and 95% confidence interval ranges from 0.67 to 0.87. For the 180⁰ meridian the correlation coefficient was 0.48, Significance level PConclusion: In terms of diagnosis of pathology and follow-up, contact lens fitting especially for the anterior cornea surface these two instruments can work quite the same that’s why a small clinical practitioner should have at least a Placido topographer. Corneal topography is an important instrument in order to measure the shape and refractive power of the cornea. Corneal topography gives practitioners an important tool to fit contact lenses asses the power and the cornea shape and helps to evaluate complications of the contact lens in relation with the corneal surface. These instruments continue to develop techniques and software in order to investigate and evaluate corneal surfaces and the integrity of the tear film, so that they can provide better solutions for their contact lens patients.Keywords
This publication has 16 references indexed in Scilit:
- Precision of 5 different keratometry devicesInternational Ophthalmology, 2015
- Variability in Oculus Pentacam Tomographer Measurements in Patients With KeratoconusCornea, 2015
- Corneal assessment technologies: Current statusSurvey of Ophthalmology, 2014
- Keratometry with five different techniques: a study of device repeatability and inter-device agreementInternational Ophthalmology, 2014
- Shape of the anterior cornea: Comparison of height data from 4 corneal topographersJournal of Cataract & Refractive Surgery, 2013
- Comparison and Repeatability of Keratometric and Corneal Power Measurements Obtained by Orbscan II, Pentacam, and Galilei Corneal Tomography SystemsAmerican Journal of Ophthalmology, 2013
- Repeatability of automatic measurements by a new Scheimpflug camera combined with Placido topographyJournal of Cataract & Refractive Surgery, 2011
- Repeatability of automatic measurements performed by a dual Scheimpflug analyzer in unoperated and post-refractive surgery eyesJournal of Cataract & Refractive Surgery, 2011
- Comparison of Pentacam and Orbscan IIz on Posterior Curvature Topography Measurements in Keratoconus EyesOphthalmology, 2006
- Corneal Topography: History, Technique, and Clinical UsesInternational Ophthalmology Clinics, 1994