Repeatability of Goldmann tonometry performed by optometry students on glaucoma patients
- 17 February 2020
- journal article
- research article
- Published by Taylor & Francis in Clinical and Experimental Optometry
- Vol. 103 (6) , 813-820
Abstract
Background The purpose of this study was to evaluate the repeatability of masked Goldmann tonometry performed by optometry students on patients with glaucoma. Methods Subjects were recruited from among patients scheduled to undergo selective laser trabeculoplasty at the Rosenberg School of Optometry clinic. Each subject had masked Goldmann tonometry performed by three examiners at each office visit: two fourth professional‐year optometry interns and an attending optometrist. Each examiner performed three sequential masked tonometry measurements on each eye. Results Twenty‐eight interns and two optometrists performed masked Goldmann tonometry on 12 glaucoma patients. The co‐efficient of variation was 9.1 per cent for the right eye and 12.1 per cent for the left eye for interns compared with 6.4 per cent right eye and 6.6 per cent left eye for optometrists. There was significant interaction between intern and patient on co‐efficient of variation (two‐factor analysis of variance, p = 0.005), indicating co‐efficient of variation was influenced by both intern and patient factors. No such interaction was found for optometrist‐performed measurements (p = 0.96). Mean interobserver difference for interns ranged between 0.9 and 3.1 mmHg, with 95 per cent limits of agreement that were proportional to mean intraocular pressure. Mean interobserver difference for optometrists ranged between 0.6 and 1.8 mmHg without proportionality bias. At higher pressure levels intern measurements became more variable and tended to overestimate optometrist measurements. Conclusions Both intraobserver and interobserver repeatability of masked tonometry was lower for interns than experienced optometrists. Intern performance differed from optometrists in that intern measurements became more variable at higher intraocular pressure levels and were significantly influenced by patient factors. The present results support the need for trainee exposure to patients with abnormally elevated intraocular pressure. Research into factors that influence trainee Goldmann tonometry repeatability is needed.Keywords
This publication has 18 references indexed in Scilit:
- Measuring intraocular pressureCurrent Opinion in Ophthalmology, 2015
- Dynamic Contour Tonometer and Goldmann Applanation Tonometer Performance in a Developing World SettingJournal of Glaucoma, 2013
- Repeatability of intraocular pressure measurements with Icare PRO rebound, Tono-Pen AVIA, and Goldmann tonometers in sitting and reclining positionsBMC Ophthalmology, 2013
- Repeatability and accuracy of applanation resonance tonometry in healthy subjects and patients with glaucomaActa Ophthalmologica, 2013
- Repeatability and Reproducibility of Goldmann Applanation, Dynamic Contour, and Ocular Response Analyzer TonometryJournal of Glaucoma, 2013
- Comparison of Intraocular Pressure Measurements and Assessment of Intraobserver and Interobserver Reproducibility With the Portable ICare Rebound Tonometer and Goldmann Applanation Tonometer in Glaucoma PatientsJournal of Glaucoma, 2013
- Intraocular Pressure Measurement Precision with the Goldmann Applanation, Dynamic Contour, and Ocular Response Analyzer TonometersOphthalmology, 2010
- A comparison of four methods of tonometry: method agreement and interobserver variabilityBritish Journal of Ophthalmology, 2005
- Reliability of intraocular pressure measurement with the Goldmann applanation tonometer in epidemiological studiesGraefe's Archive for Clinical and Experimental Ophthalmology, 1994
- THE ACCURACY OF CLINICAL APPLANATION TONOMETRYActa Ophthalmologica, 1978