Performance comparison of two automated digital morphology analyzers for leukocyte differential in patients with malignant hematological diseases: Mindray MC‐80 and Sysmex DI‐60
- 11 January 2024
- journal article
- research article
- Published by Wiley in International Journal of Laboratory Hematology
- Vol. 46 (3) , 457-465
Abstract
Background: The MC‐80 (Mindray, Shenzhen, China), a newly available artificial intelligence (AI)‐based digital morphology analyzer, is the focus of this study. We aim to compare the leukocyte differential performance of the Mindray MC‐80 with that of the Sysmex DI‐60 and the gold standard, manual microscopy.Methods: A total of 100 abnormal peripheral blood (PB) smears were compared across the MC‐80, DI‐60, and manual microscopy. Sensitivity, specificity, predictive value, and efficiency were calculated according to the Clinical and Laboratory Standards Institute (CLSI) EP12‐A2 guidelines. Comparisons were made using Bland–Altman analysis and Passing‐Bablok regression analysis. Additionally, within‐run imprecision was evaluated using five samples, each with varying percentages of mature leukocytes and blasts, in accordance with CLSI EP05‐A3 guidelines.Results: The within‐run coefficient of variation (%CV) of the MC‐80 for most cell classes in the five samples was lower than that of the DI‐60. Sensitivities for the MC‐80 ranged from 98.2% for nucleated red blood cells (NRBC) to 28.6% for reactive lymphocytes. The DI‐60's sensitivities varied between 100% for basophils and reactive lymphocytes, and 11.1% for metamyelocytes. Both analyzers demonstrated high specificity, negative predictive value, and efficiency, with over 90% for most cell classes. However, the DI‐60 showed relatively lower specificity for lymphocytes (73.2%) and lower efficiency for blasts and lymphocytes (80.1% and 78.6%, respectively) compared with the MC‐80. Bland–Altman analysis indicated that the absolute mean differences (%) ranged from 0.01 to 4.57 in MC‐80 versus manual differential and 0.01 to 3.39 in DI‐60 versus manual differential. After verification by technicians, both analyzers exhibited a very high correlation (r = 0.90–1.00) with the manual differential results in neutrophils, lymphocytes, and blasts.Conclusions: The Mindray MC‐80 demonstrated good performance for leukocyte differential in PB smears, notably exhibiting higher sensitivity for blasts identification than the DI‐60.Keywords
This publication has 19 references indexed in Scilit:
- Digital morphology analyzers in hematology: ICSH review and recommendationsInternational Journal of Laboratory Hematology, 2019
- Performance of automated digital cell imaging analyzer Sysmex DI-60Clinical Chemistry and Laboratory Medicine (CCLM), 2017
- The use of CellaVision™ DM96 in the verification of the presence of blasts in samples flagged by the Sysmex XE‐5000International Journal of Laboratory Hematology, 2017
- Daytime biological variation of hematological parameters in a healthy Chinese populationInternational Journal of Laboratory Hematology, 2016
- ICSH recommendations for the standardization of nomenclature and grading of peripheral blood cell morphological featuresInternational Journal of Laboratory Hematology, 2015
- Digital Image Analysis of Blood CellsClinics In Laboratory Medicine, 2015
- Performance evaluation of the digital cell imaging analyzer DI-60 integrated into the fully automated Sysmex XN hematology analyzer systemClinical Chemistry and Laboratory Medicine (CCLM), 2015
- Performance of CellaVision DM96 in leukocyte classificationJournal of Pathology Informatics, 2013
- Experience with CellaVision DM96 for peripheral blood differentials in a large multi-center academic hospital systemJournal of Pathology Informatics, 2012
- Can automated blood film analysis replace the manual differential? An evaluation of the CellaVision DM96 automated image analysis systemInternational Journal of Laboratory Hematology, 2009