Expiratory automatic endotracheal tube compensation reduces dynamic hyperinflation in a physical lung model
Open Access
- 23 January 2009
- journal article
- research article
- Published by Springer Nature in Critical Care
- Vol. 13 (1) , R4
Abstract
Introduction: The effect of expiratory endotracheal tube (ETT) resistance on dynamic lung inflation is unknown. We hypothesized that ETT resistance causes dynamic lung hyperinflation by impeding lung emptying. We further hypothesized that compensation for expiratory ETT resistance by automatic tube compensation (ATC) attenuates dynamic lung hyperinflation. Methods: A ventilator equipped with the original ATC mode and operating in a pressure-targeted mode was connected to a physical lung model that consists of four equally sized glass bottles filled with copper wool. Inspiratory pressure, peak expiratory flow, trapped lung volume and intrinsic positive end-expiratory pressure (PEEP) were assessed at combinations of four inner ETT diameters (7.0, 7.5, 8.0 and 8.5 mm), four respiratory rates (15, 20, 25 and 30/minute), three inspiratory pressures (3.0, 4.5 and 6.0 cmH2O) and four lung compliances (113, 86, 58 and 28 ml/cmH2O). Intrinsic PEEP was measured at the end of an expiratory hold manoeuvre. Results: At a given test lung compliance, inspiratory pressure and ETT size, increasing respiratory rates from 15 to 30/minutes had the following effects: inspiratory tidal volume and peak expiratory flow were decreased by means of 25% (range 0% to 51%) and 11% (8% to 12%), respectively; and trapped lung volume and intrinsic PEEP were increased by means of 25% (0% to 51%) and 26% (5% to 45%), respectively (all P < 0.025). At otherwise identical baseline conditions, introduction of expiratory ATC significantly attenuated (P < 0.025), by approximately 50%, the respiratory rate-dependent decreases in inspiratory tidal volume and the increases in trapped lung volume and intrinsic PEEP. Conclusions: In a lung model of pressure-targeted ventilation, expiratory ETT resistance caused dynamic lung hyperinflation during increases in respiratory rates, thereby reducing inspiratory tidal volume. Expiratory ATC attenuated these adverse effects.Keywords
This publication has 17 references indexed in Scilit:
- A new expiratory support system for resolving air trapping in lungs during mechanical ventilation: A lung model studyTechnology and Health Care, 2007
- Accuracy of automatic tube compensation in new-generation mechanical ventilators*Critical Care Medicine, 2003
- Extubation after breathing trials with automatic tube compensation, T‐tube, or pressure support ventilationActa Anaesthesiologica Scandinavica, 2002
- Ventilation with Lower Tidal Volumes as Compared with Traditional Tidal Volumes for Acute Lung Injury and the Acute Respiratory Distress SyndromeThe New England journal of medicine, 2000
- Additional inspiratory work of breathing imposed by tracheostomy tubes and non-ideal ventilator properties in critically ill patientsIntensive Care Medicine, 1999
- Breathing pattern and additional work of breathing in spontaneously breathing patients with different ventilatory demands during inspiratory pressure support and automatic tube compensationIntensive Care Medicine, 1997
- Tracheal Gas Insufflation-Pressure Control Versus Volume Control Ventilation. A Lung Model StudyAmerican Journal of Respiratory and Critical Care Medicine, 1996
- An Analysis of Desynchronization Between the Spontaneously Breathing Patient and Ventilator During Inspiratory Pressure SupportChest, 1995
- Continuous Calculation of Intratracheal Pressure in Tracheally Intubated PatientsAnesthesiology, 1993
- Determinants and limits of pressure-preset ventilation: a mathematical model of pressure controlJournal of Applied Physiology, 1989