Impact of Electrode Structure on RF-Induced Heating for an AIMD Implanted Lead in a 1.5-Tesla MRI System
- 24 April 2019
- journal article
- research article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Journal of Electromagnetics, RF and Microwaves in Medicine and Biology
- Vol. 3 (4) , 247-253
Abstract
In this paper, the impact of the electrode structure on radio frequency (RF) induced heating near an active implantable medical device (AIMD) implanted lead under magnetic resonance imaging (MRI) exposure is investigated. The specific absorption rate (SAR) distributions and the temperature rises of the leads with different electrodes were assessed. It is shown that increasing the size of lead electrode reduces the SAR distribution and temperature rise near the electrode. Our results indicate that a larger electrode size will reduce the magnitudes of the lead transfer function, and subsequently reduces the heating effect near the electrode. Both numerical simulations and experimental measurements were performed to verify the effectiveness of a large electrode in mitigating the RF-induced heating.Keywords
This publication has 20 references indexed in Scilit:
- MRI Heating Reduction for External Fixation Devices Using Absorption MaterialIEEE Transactions on Electromagnetic Compatibility, 2015
- A Technique to Evaluate MRI-Induced Electric Fields at the Ends of Practical Implanted LeadIEEE Transactions on Microwave Theory and Techniques, 2014
- Reducing RF‐related heating of cardiac pacemaker leads in MRI: Implementation and experimental verification of practical design changesMagnetic Resonance in Medicine, 2012
- Designing passive MRI‐safe implantable conducting leads with electrodesMedical Physics, 2010
- Numerical Model for Estimating RF-Induced Heating on a Pacemaker Implant During MRI: Experimental ValidationIEEE Transactions on Biomedical Engineering, 2010
- Complexity of MRI induced heating on metallic leads: Experimental measurements of 374 configurationsBioMedical Engineering OnLine, 2008
- MRI INDUCED HEATING OF DEEP BRAIN STIMULATION LEADS: EFFECT OF THE AIR-TISSUE INTERFACEProgress In Electromagnetics Research, 2008
- Calculation of MRI‐induced heating of an implanted medical lead wire with an electric field transfer functionJournal of Magnetic Resonance Imaging, 2007
- Vagus Nerve Stimulation Therapy System: In Vitro Evaluation of Magnetic Resonance Imaging-Related Heating and Function at 1.5 and 3 TeslaNeuromodulation: Technology at the Neural Interface, 2006
- A model of the temperature rise in the head due to magnetic resonance imaging proceduresMagnetic Resonance in Medicine, 1989