Evaluation of a three-dimensional ultrasound localisation system incorporating probe pressure correction for use in partial breast irradiation
Authors:
- E J HarrisEHE J Harris
- R Symonds-TaylorRSR Symonds-Taylor
- G M TreeceGTG M Treece
- A H GeeAGA H Gee
- R W PragerRPR W Prager
- P BrabantsPBP Brabants
- P M EvansPEP M Evans
- 1 October 2009
- journal article
- research article
- Published by Oxford University Press (OUP) in British Journal of Radiology
- Vol. 82 (982) , 839-846
Abstract
This work evaluates a three-dimensional (3D) freehand ultrasound-based localisation system with new probe pressure correction for use in partial breast irradiation. Accuracy and precision of absolute position measurement was measured as a function of imaging depth (ID), object depth, scanning direction and time using a water phantom containing crossed wires. To quantify the improvement in accuracy due to pressure correction, 3D scans of a breast phantom containing ball bearings were obtained with and without pressure. Ball bearing displacements were then measured with and without pressure correction. Using a single scan direction (for all imaging depths), the mean error was <1.3 mm, with the exception of the wires at 68.5 mm imaged with an ID of 85 mm, which gave a mean error of -2.3 mm. Precision was greater than 1 mm for any single scan direction. For multiple scan directions, precision was within 1.7 mm. Probe pressure corrections of between 0 mm and 2.2 mm have been observed for pressure displacements of 1.1 mm to 4.2 mm. Overall, anteroposterior position measurement accuracy increased from 2.2 mm to 1.6 mm and to 1.4 mm for the two opposing scanning directions. Precision is comparable to that reported for other commercially available ultrasound localisation systems, provided that 3D image acquisition is performed in the same scan direction. The existing temporal calibration is imperfect and a "per installation" calibration would further improve the accuracy and precision. Probe pressure correction was shown to improve the accuracy and will be useful for the localisation of the excision cavity in partial breast radiotherapy.Keywords
This publication has 21 references indexed in Scilit:
- Investigation of Accelerated Partial Breast Patient Alignment and Treatment With Helical Tomotherapy UnitInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2008
- Three-Dimensional Intrafractional Motion of Breast During Tangential Breast Irradiation Monitored With High-Sampling Frequency Using a Real-Time Tumor-Tracking Radiotherapy SystemInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2008
- Different Styles of Image-Guided RadiotherapySeminars in Radiation Oncology, 2007
- P15 The IMPORT Gold Seed Study: Evaluation of Tumour Bed Localisation and Image-guided Radiotherapy for Breast CancerClinical Oncology, 2007
- Definition of postlumpectomy tumor bed for radiotherapy boost field planning: CT versus surgical clipsInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2005
- Ultrasound probe pressure as a source of error in prostate localization for external beam radiotherapyInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2004
- Accuracy of ultrasound in localization of breast boost fieldRadiotherapy and Oncology, 2004
- Clip or Marker Migration 5–10 Weeks after Stereotactic 11-gauge Vacuum-assisted Breast Biopsy: Report of Two CasesRadiology, 2003
- Clip migration in stereotactic biopsyThe American Journal of Surgery, 2002
- Stradx: real-time acquisition and visualization of freehand three-dimensional ultrasoundMedical Image Analysis, 1999