The Impact of Size and Position of a Mechanical Expandable Transcatheter Aortic Valve: Novel Insights Through Computational Modelling and Simulation
- 23 August 2019
- journal article
- research article
- Published by Springer Nature in Journal of Cardiovascular Translational Research
- Vol. 12 (5) , 435-446
Abstract
Transcatheter aortic valve implantation has become an established procedure to treat severe aortic stenosis. Correct device sizing/positioning is crucial for optimal outcome. Lotus valve sizing is based upon multiple aortic root dimensions. Hence, it often occurs that two valve sizes can be selected. In this study, patient-specific computer simulation is adopted to evaluate the influence of Lotus size/position on paravalvular aortic regurgitation (AR) and conduction abnormalities, in patients with equivocal aortic root dimensions. First, simulation was performed in 62 patients to validate the model in terms of predicted AR and conduction abnormalities using postoperative echocardiographic, angiographic and ECG-based data. Then, two Lotus sizes were simulated at two positions in patients with equivocal aortic root dimensions. Large valve size and deep position were associated with higher contact pressure, while only large size, not position, significantly reduced the predicted AR. Despite general trends, simulations revealed that optimal device size/position is patient-specific.Keywords
This publication has 32 references indexed in Scilit:
- Assessing the impact of including leaflets in the simulation of TAVI deployment into a patient-specific aortic rootComputer Methods in Biomechanics and Biomedical Engineering, 2015
- Trends in the occurrence of new conduction abnormalities after transcatheter aortic valve implantationCatheterization and Cardiovascular Interventions, 2014
- Visualizing anatomical evidences on atrioventricular conduction system for TAVIInternational Journal of Cardiology, 2014
- Simulation of transcatheter aortic valve implantation: a patient-specific finite element approachComputer Methods in Biomechanics and Biomedical Engineering, 2012
- Prosthesis Oversizing in Balloon-Expandable Transcatheter Aortic Valve Implantation Is Associated With Contained Rupture of the Aortic RootCirculation: Cardiovascular Interventions, 2012
- Patient-specific simulations of transcatheter aortic valve stent implantationMedical & Biological Engineering & Computing, 2012
- Transcatheter versus Surgical Aortic-Valve Replacement in High-Risk PatientsNew England Journal of Medicine, 2011
- Gross anatomy of the human cardiac conduction system with comparative morphological and developmental implications for human applicationAnnals of Anatomy - Anatomischer Anzeiger, 2011
- Anatomic and Procedural Predictors of Paravalvular Aortic Regurgitation After Implantation of the Medtronic CoreValve BioprosthesisJACC, 2010
- Left retrograde cardioangiography in acquired cardiac diseaseThe American Journal of Cardiology, 1964