Evaluation of the feasibility of a multisource CBCT for maxillofacial imaging
- 14 August 2023
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 68 (17) , 175012
Abstract
Objective . The aim of this study was to investigate the feasibility of improving the image quality and accuracy of cone beam computed tomography (CBCT) by replacing the conventional wide cone angle x-ray tube with a distributed x-ray source array positioned in the axial direction. Approach . The multisource CBCT (ms-CBCT) design was experimentally simulated using a benchtop scanner with a carbon nanotube x-ray tube and a flat-panel detector. The source was collimated and translated in the axial direction to simulate a source array with a reduced cone angle for each beam. An adjacent scatter ratio subtraction (ASRS) method was implemented for residual scatter reduction. Several phantoms were imaged using the ms-CBCT and conventional CBCT configurations under otherwise similar conditions. The Requirements of the ms-CBCT design on the x-ray source and detector were evaluated. Main results . Compared to the conventional CBCT, the ms-CBCT design with 8 sources and ASRS significantly improved the image quality and accuracy, including: (1) reducing the cupping artifact from 15% to 3.5%; (2) reducing the spatial nonuniformity of the CT Hounsfield unit values from 38.0 to 9.2; (3) improving the contrast-to-noise ratio of the low contrast objects (acrylic and low density polyethylene inserts) against the water-equivalent background by ∼20% and (4) reducing the root-mean-square error of the HU values by 70%, from 420.1 to 124.4. The imaging dose and scanning time used by the current clinical CBCT for maxillofacial imaging can be achieved by current source and detector technologies. Significance . The ms-CBCT design significantly reduces the scatter and improves the image quality and accuracy compared to the conventional CBCT.Keywords
Funding Information
- National Institutes of Health (R56DE030962-01)
- North Carolina Biotechnology Center (2021-TRG-6725)
This publication has 42 references indexed in Scilit:
- X‐ray scatter correction method for dedicated breast computed tomographyMedical Physics, 2012
- Artefacts in CBCT: a reviewDentomaxillofacial Radiology, 2011
- Tetrahedron beam computed tomography (TBCT): a new design of volumetric CT systemPhysics in Medicine & Biology, 2009
- 3D Analytic Cone‐Beam Reconstruction for Multiaxial CT AcquisitionsInternational Journal of Biomedical Imaging, 2009
- What is Cone-Beam CT and How Does it Work?Dental Clinics of North America, 2008
- Cone-Beam CT Diagnostic Applications: Caries, Periodontal Bone Assessment, and Endodontic ApplicationsDental Clinics of North America, 2008
- Efficiency of antiscatter grids for flat-detector CTPhysics in Medicine & Biology, 2007
- A prototype table-top inverse-geometry volumetric CT systemMedical Physics, 2006
- Stationary scanning x-ray source based on carbon nanotube field emittersApplied Physics Letters, 2005
- X‐ray micro‐CT with a displaced detector arrayMedical Physics, 2002