Automation of cellular therapy product manufacturing: results of a split validation comparing CD34 selection of peripheral blood stem cell apheresis product with a semi-manual vs. an automatic procedure
Authors:
- Christiane HümmerCHChristiane Hümmer
- Carolin PoppeCPCarolin Poppe
- Milica BunosMBMilica Bunos
- Belinda StockBSBelinda Stock
- Eva WingenfeldEWEva Wingenfeld
- Volker HuppertVHVolker Huppert
- Juliane StuthJSJuliane Stuth
- Kristina ReckKRKristina Reck
- Mike EsslMEMike Essl
- Erhard SeifriedESErhard Seifried
- Halvard BonigHBHalvard Bonig
Open Access
- 16 March 2016
- journal article
- research article
- Published by Springer Nature in Journal of Translational Medicine
- Vol. 14 (1) , 1-7
Abstract
Automation of cell therapy manufacturing promises higher productivity of cell factories, more economical use of highly-trained (and costly) manufacturing staff, facilitation of processes requiring manufacturing steps at inconvenient hours, improved consistency of processing steps and other benefits. One of the most broadly disseminated engineered cell therapy products is immunomagnetically selected CD34+ hematopoietic “stem” cells (HSCs). As the clinical GMP-compliant automat CliniMACS Prodigy is being programmed to perform ever more complex sequential manufacturing steps, we developed a CD34+ selection module for comparison with the standard semi-automatic CD34 “normal scale” selection process on CliniMACS Plus, applicable for 600 × 106 target cells out of 60 × 109 total cells. Three split-validation processings with healthy donor G-CSF-mobilized apheresis products were performed; feasibility, time consumption and product quality were assessed. All processes proceeded uneventfully. Prodigy runs took about 1 h longer than CliniMACS Plus runs, albeit with markedly less hands-on operator time and therefore also suitable for less experienced operators. Recovery of target cells was the same for both technologies. Although impurities, specifically T- and B-cells, were 5 ± 1.6-fold and 4 ± 0.4-fold higher in the Prodigy products (p = ns and p = 0.013 for T and B cell depletion, respectively), T cell contents per kg of a virtual recipient receiving 4 × 106 CD34+ cells/kg was below 10 × 103/kg even in the worst Prodigy product and thus more than fivefold below the specification of CD34+ selected mismatched-donor stem cell products. The products’ theoretical clinical usability is thus confirmed. This split validation exercise of a relatively short and simple process exemplifies the potential of automatic cell manufacturing. Automation will further gain in attractiveness when applied to more complex processes, requiring frequent interventions or handling at unfavourable working hours, such as re-targeting of T-cells.Keywords
This publication has 12 references indexed in Scilit:
- Automated isolation of primary antigen‐specific T cells from donor lymphocyte concentrates: results of a feasibility exerciseVox Sanguinis, 2015
- Red blood cell depletion from bone marrow and peripheral blood buffy coat: a comparison of two new and three established technologiesTransfusion, 2015
- Deferrals of volunteer stem cell donors referred for evaluation for matched-unrelated stem cell donationBone Marrow Transplantation, 2014
- Unstimulated leukapheresis in patients and donors: comparison of two apheresis systemsTransfusion, 2013
- Integrated Clinical Scale Manufacturing System for Cellular Products Derived by Magnetic Cell Separation, Centrifugation and Cell CultureChemie Ingenieur Technik, 2013
- Allogeneic donor peripheral blood “stem cell” apheresis: prospective comparison of two apheresis systemsTransfusion, 2011
- Mobilized allogeneic peripheral stem/progenitor cell apheresis with Spectra Optia v.5·0, a novel, automatic interface‐controlled apheresis system: results from the first feasibility trialVox Sanguinis, 2011
- Enumeration of viable CD34+ cells by flow cytometry in blood, bone marrow and cord blood: results of a study of the novel BD™ stem cell enumeration kitCytotherapy, 2011
- Isolation and Enrichment of Stem Cells2009
- Isolation of Highly Purified Autologous and Allogeneic Peripheral CD34+ Cells Using the CliniMACS DeviceJournal of Hematotherapy, 1999