Bandwidth and Common Mode Optimization for Current and Voltage Sources in Bioimpedance Spectroscopy
Open Access
- 1 January 2021
- journal article
- research article
- Published by Walter de Gruyter GmbH in Journal of Electrical Bioimpedance
- Vol. 12 (1) , 135-146
Abstract
Bioimpedance measurements use current or voltage sources to inject an excitation signal into the body. These sources require a high bandwidth, typically from 1 kHz to 1 MHz. Besides a low common mode, current limitation is necessary for patient safety. In this paper, we compare a symmetric enhanced Howland current source (EHCS) and a symmetric voltage source (VS) based on a non-inverting amplifier between 1 kHz and 1 MHz. A common mode reduction circuit has been implemented in both sources. The bandwidth of each source was optimized in simulations and achieved a stable output impedance over the whole frequency range. In laboratory measurements, the output impedance of the EHCS had its -3 dB point at 400 kHz. In contrast, the VS reached the +3 dB point at 600 kHz. On average over the observed frequency range, the active common mode compensation achieved a common mode rejection of -57.7 dB and -71.8 dB for the EHCS and VS, respectively. Our modifications to classical EHCS and VS circuits achieved a low common mode signal between 1 kHz and 1 MHz without the addition of complex circuitry, like general impedance converters. As a conclusion we found VSs to be superior to EHCSs for bioimpedance spectroscopy due to the higher bandwidth performance. However, this only applies if the injected current of the VS can be measured.Keywords
This publication has 22 references indexed in Scilit:
- Electrode-Skin contact impedance: In vivo measurements on an ovine modelJournal of Physics: Conference Series, 2013
- Bio-impedance Excitation System: A Comparison of Voltage Source and Current Source DesignsAPCBEE Procedia, 2013
- Performance evaluation of wideband bio-impedance spectroscopy using constant voltage source and constant current sourceMeasurement Science and Technology, 2012
- Front end with offset-free symmetrical current source optimized for time domain impedance spectroscopyPhysiological Measurement, 2011
- Comparison of Howland and General Impedance Converter (GIC) circuit based current sources for bio-impedance measurementsJournal of Physics: Conference Series, 2010
- Comparison of a new integrated current source with the modified Howland circuit for EIT applicationsPhysiological Measurement, 2009
- Low-noise current excitation sub-system for medical EITPhysiological Measurement, 2008
- Complex bioelectric impedance measurement system for the frequency range from 5 Hz to 1 MHzAnnals of Biomedical Engineering, 1993
- The Sheffield data collection systemClinical Physics and Physiological Measurement, 1987
- Electrical Properties of Tissue and Cell Suspensions1957