Advances in Spiral fMRI: A High-resolution Study with Single-shot Acquisition
Authors:
- Lars KasperLKLars Kasper
- Maria EngelMEMaria Engel
- Jakob HeinzleJHJakob Heinzle
- Matthias Mueller-SchraderMMMatthias Mueller-Schrader
- Nadine N. GraedelNGNadine N. Graedel
- Jonas ReberJRJonas Reber
- Thomas SchmidTSThomas Schmid
- Christoph BarmetCBChristoph Barmet
- Bertram Jakob WilmBWBertram Jakob Wilm
- Klaas Enno StephanKSKlaas Enno Stephan
- Klaas Paul PruessmannKPKlaas Paul Pruessmann
Preprint
Abstract
Spiral fMRI has been put forward as a viable alternative to rectilinear echo-planar imaging, in particular due to its enhanced average k-space speed and thus high acquisition efficiency. This renders spirals attractive for contemporary fMRI applications that require high spatiotemporal resolution, such as laminar or columnar fMRI. However, in practice, spiral fMRI is typically hampered by its reduced robustness and ensuing blurring artifacts, which arise from imperfections in both static and dynamic magnetic fields. Recently, these limitations have been overcome by the concerted application of an expanded signal model that accounts for such field imperfections, and its inversion by iterative image reconstruction. In the challenging ultra-high field environment of 7 Tesla, where field inhomogeneity effects are aggravated, both multi-shot and single-shot 2D spiral imaging at sub-millimeter resolution was demonstrated with high depiction quality and anatomical congruency. In this work, we further these advances towards a time series application of spiral readouts, namely, single-shot spiral BOLD fMRI at 0.8 mm in-plane resolution. We demonstrate that high-resolution spiral fMRI at 7 T is not only feasible, but delivers both excellent image quality, BOLD sensitivity, and spatial specificity of the activation maps, with little artifactual blurring. Furthermore, we show the versatility of the approach with a combined in/out spiral readout at a more typical resolution (1.5 mm), where the high acquisition efficiency allows to acquire two images per shot for improved sensitivity by echo combination. Highlights: This work reports the first fMRI study at 7T with high-resolution spiral readout gradient waveforms. We achieve spiral fMRI with sub-millimeter resolution (0.8 mm, FOV 230 mm), acquired in a single shot (36 slices in 3.3 s). Spiral images exhibit intrinsic geometric congruency to anatomical scans, and spatially specific activation patterns. Image reconstruction rests on a signal model expanded by measured trajectories and static field maps, inverted by cg-SENSE. We assess generalizability of the approach for spiral in/out readouts, providing two images per shot (1.5 mm resolution).Keywords
All Related Versions
- Published version: NeuroImage, 246, 118738.
This publication has 158 references indexed in Scilit:
- Differentiating BOLD and non-BOLD signals in fMRI time series using multi-echo EPINeuroImage, 2012
- Cytoarchitectonic mapping of the human dorsal extrastriate cortexBrain Structure and Function, 2012
- Spurious but systematic correlations in functional connectivity MRI networks arise from subject motionNeuroImage, 2012
- Applications of arterial spin labeled MRI in the brainJournal of Magnetic Resonance Imaging, 2012
- Blipped‐controlled aliasing in parallel imaging for simultaneous multislice echo planar imaging with reduced g‐factor penaltyMagnetic Resonance in Medicine, 2011
- Dynamic B0 shimming at 7 TMagnetic Resonance Imaging, 2011
- SPIRiT: Iterative self‐consistent parallel imaging reconstruction from arbitrary k‐spaceMagnetic Resonance in Medicine, 2010
- Three dimensional echo-planar imaging at 7 TeslaNeuroImage, 2010
- Interleaved spiral‐in/out with application to functional MRI (fMRI)Magnetic Resonance in Medicine, 2009
- Joint estimation of water/fat images and field inhomogeneity mapMagnetic Resonance in Medicine, 2008