Dosimetric evaluation in Helical TomoTherapy for lung SBRT using Monte Carlo‐based independent dose verification software
Open Access
- 18 February 2024
- journal article
- research article
- Published by Wiley in Journal of Applied Clinical Medical Physics
- Vol. 25 (5) , e14305
Abstract
Purpose: To elucidate the dosimetric errors caused by a model‐based algorithm in lung stereotactic body radiation therapy (SBRT) with Helical TomoTherapy (HT) using Monte Carlo (MC)‐based dose verification software.Methods: For 38 plans of lung SBRT, the dose calculation accuracy of a treatment planning system (TPS) of HT was compared with the results of DoseCHECK, the commercial MC‐based independent verification software. The following indices were extracted to evaluate the correlation of dosimetric errors: (1) target volume, (2) average computed tomography (CT) value of the planning target volume (PTV) margin, and (3) average CT value of surrounding 2‐mm area of the PTV (PTV ring). Receiver operating characteristic (ROC) analyses determined the threshold for 5% of differences in PTV D95%. Then, the 38 plans were classified into two groups using the cutoff values of ROC analysis for these three indices. Dosimetric differences between groups were statistically compared using the Mann–Whitney U test.Results: TPS of HT overestimated by more than 5% in the PTV D95% in 16 of 38 plans. The PTV ring showed the strongest correlation with dosimetric differences. The cutoff value for the target volume, the PTV margin, and the PTV ring was 14.7 cc, −754 HU, and −708 HU, respectively. The area under the curve (AUC) for the target volume, the PTV margin, and the PTV ring were 0.835, 0.878, and 0.932, respectively. Dosimetric errors more than 5% were observed when the PTV volume was less than 15 cc or when the CT value around the target was less than −700 HU.Conclusion: The TPS of HT might overestimate the PTV dose by more than 5% if any the three indices in this study were below threshold. Therefore, independent verification with an MC‐based algorithm should be strongly recommended for lung SBRT in HT.Keywords
This publication has 36 references indexed in Scilit:
- Sensitivity and specificity of secondary dose calculation for head and neck treatment plansJournal of Applied Clinical Medical Physics, 2023
- A structured FMEA approach to optimizing combinations of plan‐specific quality assurance techniques for IMRT and VMAT QAMedical Physics, 2023
- Optimal threshold of a control parameter for tomotherapy respiratory tracking: A phantom studyJournal of Applied Clinical Medical Physics, 2023
- A hybrid method to improve efficiency of patient specific SRS and SBRT QA using 3D secondary dose verificationJournal of Applied Clinical Medical Physics, 2022
- Commissioning and routine quality assurance of the Radixact Synchrony systemMedical Physics, 2021
- Report of AAPM Task Group 219 on independent calculation‐based dose/MU verification for IMRTMedical Physics, 2021
- Evaluation of radixact motion synchrony for 3D respiratory motion: Modeling accuracy and dosimetric fidelityJournal of Applied Clinical Medical Physics, 2020
- Independent calculation‐based verification of volumetric‐modulated arc therapy–stereotactic body radiotherapy plans for lung cancerJournal of Applied Clinical Medical Physics, 2020
- Validation of the Acuros XB dose calculation algorithm versus Monte Carlo for clinical treatment plansMedical Physics, 2018
- Tolerance limits and methodologies forIMRTmeasurement‐based verificationQA:Recommendations ofAAPMTask Group No. 218Medical Physics, 2018