Comparative analysis of hemoglobin a1c measurement methods: inter-device agreement and clinical impact
Authors:
Open Access
- 4 June 2026
- journal article
- Published by Springer Nature in Scientific Reports
- Vol. 16 (1)
Abstract
Glycated hemoglobin (HbA1c) is widely recommended for the diagnosis and monitoring of diabetes mellitus (DM). Minimal changes in HbA1c concentration can influence clinical decisions, making accurate and interference-free measurement essential. To compare the analytical performance and inter-method agreement of three HbA1c measurement techniques-Capillary Electrophoresis, High-Performance Liquid Chromatography (HPLC), and Turbidimetric Immunoinhibition-used in routine clinical practice. Fifty EDTA-anticoagulated patient samples were consecutively analyzed using HPLC (Tosoh G8), Capillary Electrophoresis (Sebia Capillarys 3 Tera), and an autoanalyzer (Beckman DXC 700 AU). Results were stratified into four HbA1c ranges (4-6%, 6-8%, 8-10%, 10-12%) and analyzed using Pearson correlation, Bland-Altman, and Passing-Bablok regression. Capillary Electrophoresis and HPLC showed a high degree of agreement (r = 0.998, bias = 0.07). Capillary Electrophoresis vs. autoanalyzer also demonstrated high correlation (r = 0.997) but a larger bias (0.52) that increased at higher concentrations. HPLC vs. autoanalyzer yielded r = 0.997 and bias = 0.45, with proportional bias at high HbA1c levels. Passing-Bablok analysis confirmed concentration-dependent deviations for the autoanalyzer. Capillary Electrophoresis and HPLC provide highly consistent HbA1c results, while the autoanalyzer may introduce clinically relevant bias, particularly at higher HbA1c levels, which may influence clinical interpretation near commonly used decision thresholds. However, given the limited sample size, these findings should be interpreted with caution.Keywords
This publication has 13 references indexed in Scilit:
- Analytical performance evaluation of the Mindray enzymatic assay for hemoglobin A1c measurementScientific Reports, 2024
- Interference of hemoglobin variants with HbA1c measurements by six commonly used HbA1c methodsLaboratory Medicine, 2024
- Role of liver parameters in diabetes mellitus – a narrative reviewEndocrine Regulations, 2023
- Interference of hemoglobin variants in HbA1c quantificationClinica Chimica Acta, 2022
- Differential Effects on HbA1c Detection by HPLC and Capillary Electrophoresis in Five Types of Hb Variants in ChinaLaboratory Medicine, 2022
- Glycated Hemoglobin Measurement: Comparison of Three Methods Versus High Performance Liquid ChromatographyJournal of Diabetes Science and Technology, 2021
- Comparison of Turbidimetric Inhibition Immunoassay, High-Performance Liquid Chromatography, and Capillary Electrophoresis Methods for Glycated Hemoglobin DeterminationLaboratory Medicine, 2020
- Global aetiology and epidemiology of type 2 diabetes mellitus and its complicationsNature Reviews Endocrinology, 2017
- HbA1c analysis by capillary electrophoresis – comparison with chromatography and an immunological methodScandinavian Journal of Clinical and Laboratory Investigation, 2017
- Performances of Capillary Electrophoresis and HPLC Methods in HbA1c Determination: Diagnostic Accuracy in HbS and HbD‐Iran Variants' PresenceJournal of Clinical Laboratory Analysis, 2014