Validation of scatter limitation correction to eliminate scatter correction error in oxygen-15 gas-inhalation positron emission tomography images
Authors:
- 1 October 2018
- journal article
- research article
- Published by Wolters Kluwer Health in Nuclear Medicine Communications
- Vol. 39 (10) , 936-944
Abstract
High levels of radioactivity inside a facemask cause scatter correction (SC) errors that appear as photopenic artifacts on quantitative oxygen-15 (15O) gas-inhalation positron emission tomography (PET) images. The present study aimed to validate the ability of scatter limitation correction (SLC) to eliminate SC errors in 15O gas-inhalation PET images acquired from patients and a phantom. We analyzed the SC errors in phantom images and calculated parametric images of the cerebral blood flow (CBF), cerebral blood volume, oxygen extraction fraction (OEF), and cerebral metabolic rate of oxygen (CMRO2). Phantoms comprised a cylinder and paper with radioactivity to simulate a facemask during (15O)O2 gas inhalation. Parametric images were calculated from 15O gas-inhalation PET images of ten participants. All PET data were reconstructed using conventional SC as model-based SC and SLC. Images acquired from the phantoms and parametric images were assessed visually and quantitatively in the presence and absence of SC error. SC error was evident in images derived from the paper phantom and at the slice level of the cerebellum in CBF, OEF, and CMRO2 images. The radioactivity concentration in the cylindrical phantom with the paper phantom significantly improved with SLC. The SLC also increased the quantitative indices of CBF, OEF, and CMRO2 by 23.8, 42.2, and 44.4%, respectively. SLC visually eliminated the SC error and increased the quantitative parameters on 15O gas-inhalation images derived from a phantom and from patients.Keywords
This publication has 16 references indexed in Scilit:
- Effects of a novel tungsten-impregnated rubber neck shield on the quality of cerebral images acquired using 15O-labeled gasRadiological Physics and Technology, 2017
- Validity of using a 3-dimensional PET scanner during inhalation of15O-labeled oxygen for quantitative assessment of regional metabolic rate of oxygen in manPhysics in Medicine & Biology, 2014
- Detection of misery perfusion in the cerebral hemisphere with chronic unilateral major cerebral artery steno-occlusive disease using crossed cerebellar hypoperfusion: comparison of brain SPECT and PET imagingEuropean Journal of Nuclear Medicine and Molecular Imaging, 2013
- Effect of Patient Arm Motion in Whole-Body PET/CTJournal of Nuclear Medicine, 2011
- The effect of activity outside the field-of-view on image signal-to-noise ratio for 3D PET with15OPhysics in Medicine & Biology, 2011
- Evaluation of dynamic row-action maximum likelihood algorithm reconstruction for quantitative 15O brain PETAnnals of Nuclear Medicine, 2009
- Optimal scan time of oxygen-15-labeled gas inhalation autoradiographic method for measurement of cerebral oxygen extraction fraction and cerebral oxygen metabolic rateAnnals of Nuclear Medicine, 2008
- CT-based attenuation correction in the calculation of semi-quantitative indices of [18F]FDG uptake in PETEuropean Journal of Nuclear Medicine and Molecular Imaging, 2002
- Model-based scatter correction for fully 3D PETPhysics in Medicine & Biology, 1996
- Reversal of focal "misery-perfusion syndrome" by extra-intracranial arterial bypass in hemodynamic cerebral ischemia. A case study with 15O positron emission tomography.Stroke, 1981