Abstract
Plethysmographic measurement of airway resistance (Raw) has been determined by single-point analysis, usually at 50% of maximum inspiratory flow (MIF). Computer-assisted (best fit) analysis, however, allows Raw to be calculated by applying a regression line to any portion of the plethysmograph pressure–flow loop. We determined whether the results of best fit analysis using a computer program, sampling at 200 Hz, were influenced by the portion of the inspiratory loop analysed and if best fit or single-point analysis gave more reproducible results. Twenty infants of median gestational age 26 (range 24–28) weeks, were studied at a median age of 12 (12–14) months corrected for prematurity. Raw was calculated by best fit analysis between 0 and 33% MIF, 0 and 50% MIF and 0 and 67% MIF and single-point analysis at 50% of MIF. Similar mean Raw values were obtained by best fit analysis between 0 and 33% MIF (2.79 kPa/(l/s)) and 0 and 50% MIF (3.01 kPa/(l/s)) and single-point analysis at 50% MIF (2.86 kPa/(l/s)), but best fit analysis between 0 and 67% gave higher results (3.60 kPa/(l/s)), p < 0.0001. Within the linear portion of the inspiratory loop, the mean intrasubject coefficient of variation was lowest for best fit analysis between 0 and 50% MIF. Best fit computerized analysis between 0 and 50% MIF is recommended as the analysis of choice.