Implementation and evaluation of the AAPM TG 111 for radiation dosimetry in a state-of-the-art C-arm cone beam CT system
- 1 September 2023
- journal article
- research article
- Published by IOP Publishing in Journal of Radiological Protection
- Vol. 43 (3) , 031521
Abstract
The aim of this study was to assess the dosimetric characteristics of a state-of-the-art C-arm cone beam computed tomography (CBCT) system using the methodology proposed by the American Association of Physicists in Medicine (AAPM) Task Group (TG) 111. The dose measurement methodology described in AAPM TG 111 for wide cone beam acquisitions without table translation was employed to estimate equilibrium beam length ( α eq ‾ ) and equilibrium dose ( f eq ‾ ) in various interventional task-specific protocols with different tube arc projection geometries. Dose profiles were derived from point dose measurements in the centre and peripheral locations of the ICRU/AAPM and standard polymethyl methacrylate (PMMA) body phantom. Dose measurements were performed in phantom and free-in-air using a solid-state point detector. Monte Carlo (MC) based simulation dosimetry was performed to quantify the inhomogeneous dose patterns imparted in the phantoms. Estimated α eq ‾ and f eq ‾ on the ICRU/AAPM phantom was up to 49.4 cm and 6.17 mGy/100 mAs, respectively. Corresponding values determined on the PMMA phantom were 139 cm and 8.8 mGy/100 mAs, respectively. Free-in-air dose measurement ranged from 1.43 mGy/100 mAs to 5.93 mGy/100 mAs. Per cent difference in α eq ‾ and f eq ‾ between MC simulation and solid-state point detector measurement methods in the ICRU/AAPM phantom were within 16% and 18%, respectively. Manufacturers can use the presented methodology to characterize the dosimetric properties of C-arm CBCT systems. Clinical medical physicists may follow this methodology to verify corresponding data provided by the manufacturer and check for C-arm CBCT system performance dosimetric consistency.Keywords
This publication has 13 references indexed in Scilit:
- Assessment of abdominal organ dose and image quality in varying arc trajectory interventional C-arm cone beam CTPhysica Medica, 2022
- Organ doses and normalized organ doses for various age groups in ultralow dose pediatric C-arm cone-beam CTEuropean Radiology, 2022
- A comprehensive Monte Carlo study of CT dose metrics proposed by the AAPM Reports 111 and 200Medical Physics, 2021
- Development of a method to estimate organ doses for pediatric CT examinationsMedical Physics, 2016
- Monte Carlo assessment of CT dose equilibration in PMMA and water cylinders with diameters from 6 to 55 cmMedical Physics, 2013
- Estimation of the weighted CTDI∞for multislice CT examinationsMedical Physics, 2012
- A method of estimating conceptus doses resulting from multidetector CT examinations during all stages of gestationMedical Physics, 2010
- Cone beam CT dosimetry: A unified and self‐consistent approach including all scan modalities—With or without phantom motionMedical Physics, 2010
- Validation of a Monte Carlo tool for patient-specific dose simulations in multi-slice computed tomographyEuropean Radiology, 2007
- Enlarged longitudinal dose profiles in cone‐beam CT and the need for modified dosimetryMedical Physics, 2005