Influence of the characteristic curve on the clinical image quality of lumbar spine and chest radiographs
Authors:
- A TingbergATA Tingberg
- C HerrmannCHC Herrmann
- B LanhedeBLB Lanhede
- A AlménAAA Almén
- M SandborgMSM Sandborg
- G McVeyGMG McVey
- S MattssonSMS Mattsson
- W PanzerWPW Panzer
- J BesjakovJBJ Besjakov
- L G MånssonLML G Månsson
- S KheddacheSKS Kheddache
- G Alm CarlssonGCG Alm Carlsson
- 1 March 2004
- journal article
- Published by Oxford University Press (OUP) in British Journal of Radiology
- Vol. 77 (915) , 204-215
Abstract
The "European Guidelines on Quality Criteria for Diagnostic Radiographic Images" do not address the choice of the film characteristic (H&D) curve, which is an important parameter for the description of a radiographic screen-film system. The image contrast of clinical lumbar spine and chest radiographs was altered by digital image processing techniques, simulating images with different H&D curves, both steeper and flatter than the original. The manipulated images were printed on film for evaluation. Seven experienced radiologists evaluated the clinical image quality by analysing the fulfilment of the European Image Criteria (ICS) and by visual grading analysis (VGA) of in total 224 lumbar spine and 360 chest images. A parallel study of the effect of the H&D curve has also been made using a theoretical model. The contrast (DeltaOD) of relevant anatomical details was calculated, using a Monte Carlo simulation-model of the complete imaging system including a 3D voxel phantom of a patient. Correlations between the calculated contrast and the radiologists' assessment by VGA were sought. The results of the radiologists' assessment show that the quality in selected regions of lumbar spine and chest images can be significantly improved by the use of films with a steeper H&D curve compared with the standard latitude film. Significant (p<0.05) correlations were found between the VGA results and the calculations of the contrast of transverse processes and trabecular details in the lumbar spine vertebrae, and with the contrast of blood vessels in the retrocardiac area of the chest.Keywords
This publication has 17 references indexed in Scilit:
- A study and optimization of lumbar spine X-ray imaging systemsBritish Journal of Radiology, 2003
- The influence of different technique factors on image quality of chest radiographs as evaluated by modified CEC image quality criteriaBritish Journal of Radiology, 2002
- Demonstration of correlations between clinical and physical image quality measures in chest and lumbar spine screen–film radiographyBritish Journal of Radiology, 2001
- The influence of different technique factors on image quality of lumbar spine radiographs as evaluated by established CEC image criteria.British Journal of Radiology, 2000
- The Influence of Different Technique Factors on Image Quality for Chest Radiographys: Application of the Recent CEC Image Quality CriteriaRadiation Protection Dosimetry, 2000
- A Voxel Phantom Based Monte Carlo Computer Program for Optimisation of Chest and Lumbar Spine X Ray Imaging SystemsRadiation Protection Dosimetry, 2000
- Free-response methodology: alternate analysis and a new observer-performance experiment.Radiology, 1990
- ROC Methodology in Radiologic ImagingInvestigative Radiology, 1986
- Detection accuracy in chest radiographyAmerican Journal of Roentgenology, 1978
- Factors Affecting the Detectability of a Simulated Radiographic SignalInvestigative Radiology, 1973