High-sensitivity CMUT with interlayer metal architecture for medical ultrasonic systems
Authors:
- 1 June 2015
- conference paper
- Published by Institute of Electrical and Electronics Engineers (IEEE)
- No. 2159547X,p. 1211-1214
Abstract
This paper presents an integration of the capacitive micromachined ultrasonic transducers (CMUTs) and the transimpedance amplifier in medical ultrasound systems. The proposed CMUT and the sensing circuits are both fabricated on the same chip in a 0.18μm CMOS MEMS technology. Two interlayer metal structures are proposed to improve the sensitivity. The CMUTs are designed with 2MHz cells for medical diagnosis. The conversion efficiency of the CMUTs is 4 times and 1.8 times higher than that in the published structures.Keywords
This publication has 5 references indexed in Scilit:
- Front-end receiver electronics for high-frequency monolithic CMUT-on-CMOS imaging arraysIEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 2011
- Design and characterization of the immersion-type capacitive ultrasonic sensors fabricated in a CMOS processJournal of Micromechanics and Microengineering, 2011
- Modeling and Characterization of CMOS-Fabricated Capacitive Micromachined Ultrasound TransducersJournal of Microelectromechanical Systems, 2010
- CMUT-in-CMOS ultrasonic transducer arrays with on-chip electronics2009
- Biocompatible coatings for CMUTs in a harsh, aqueous environmentJournal of Micromechanics and Microengineering, 2007