Radiation dose reduction to the breast in thoracic CT: Comparison of bismuth shielding, organ‐based tube current modulation, and use of a globally decreased tube current
- 21 October 2011
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 38 (11) , 6084-6092
Abstract
Purpose:: The purpose of this work was to evaluate dose performance and image quality in thoracic CT using three techniques to reduce dose to the breast: bismuth shielding, organ‐based tube current modulation (TCM) and global tube current reduction.Methods:: Semi‐anthropomorphic thorax phantoms of four different sizes (15, 30, 35, and 40 cm lateral width) were used for dose measurement and image quality assessment. Four scans were performed on each phantom using 100 or 120 kV with a clinical CT scanner: (1) reference scan; (2) scan with bismuth breast shield of an appropriate thickness; (3) scan with organ‐based TCM; and (4) scan with a global reduction in tube current chosen to match the dose reduction from bismuth shielding. Dose to the breast was measured with an ion chamber on the surface of the phantom. Image quality was evaluated by measuring the mean and standard deviation of CT numbers within the lung and heart regions.Results:: Compared to the reference scan, dose to the breast region was decreased by about 21% for the 15‐cm phantom with a pediatric (2‐ply) shield and by about 37% for the 30, 35, and 40‐cm phantoms with adult (4‐ply) shields. Organ‐based TCM decreased the dose by 12% for the 15‐cm phantom, and 34–39% for the 30, 35, and 40‐cm phantoms. Global lowering of the tube current reduced breast dose by 23% for the 15‐cm phantom and 39% for the 30, 35, and 40‐cm phantoms. In phantoms of all four sizes, image noise was increased in both the lung and heart regions with bismuth shielding. No significant increase in noise was observed with organ‐based TCM. Decreasing tube current globally led to similar noise increases as bismuth shielding. Streak and beam hardening artifacts, and a resulting artifactual increase in CT numbers, were observed for scans with bismuth shields, but not for organ‐based TCM or global tube current reduction.Conclusions:: Organ‐based TCM produces dose reduction to the breast similar to that achieved with bismuth shielding for both pediatric and adult phantoms. However, organ‐based TCM does not affect image noise or CT number accuracy, both of which are adversely affected by bismuth shielding. Alternatively, globally decreasing the tube current can produce the same dose reduction to the breast as bismuth shielding, with a similar noise increase, yet without the streak artifacts and CT number errors caused by the bismuth shields. Moreover, globally decreasing the tube current reduces the dose to all tissues scanned, not simply to the breast.Keywords
This publication has 21 references indexed in Scilit:
- Dose reduction in multislice CT by means of bismuth shields: results of in vivo measurements and computed evaluationLa radiologia medica, 2009
- Radiation Dose to the Thyroid Gland and Breast From Multidetector Computed Tomography of the Cervical SpineJournal of Computer Assisted Tomography, 2009
- Exposure to Low-Dose Ionizing Radiation from Medical Imaging ProceduresNew England Journal of Medicine, 2009
- Thyroid Shields versus z-Axis Automatic Tube Current Modulation for Dose Reduction at Neck CTRadiology, 2008
- Computed Tomography — An Increasing Source of Radiation ExposureThe New England journal of medicine, 2007
- MOSFET Dosimetry for Radiation Dose Assessment of Bismuth Shielding of the Eye in ChildrenAmerican Journal of Roentgenology, 2007
- Female Breast Radiation Exposure During Thorax Multidetector Computed Tomography and the Effectiveness of Bismuth Breast Shield to Reduce Breast Radiation DoseJournal of Computer Assisted Tomography, 2007
- Radiation dose reduction to breast and thyroid during MDCT: effectiveness of an in-plane bismuth shieldActa Radiologica, 2006
- Orbital, thyroid, and breast superficial radiation shielding for patients undergoing diagnostic CTSeminars in Ultrasound, CT and MRI, 2002
- The breast: in-plane x-ray protection during diagnostic thoracic CT--shielding with bismuth radioprotective garments.Radiology, 1997