Top Cited Papers
Open Access
Abstract
We report on a 88 Sr + single-ion optical clock with an estimated fractional systematic uncertainty of 7.9× 10 − 19 . The low uncertainty is enabled by small rf losses, a thorough evaluation of the blackbody-radiation temperature, and our recent measurement of the differential polarizability. A detailed uncertainty evaluation is presented. We also report on two absolute frequency measurements: one against a remote cesium fountain clock, and one against International Atomic Time (TAI). The former lasted 12 d and resulted in a frequency value of 444 779 044 095 485.49(15) Hz. The latter spanned 10 months with monthly optical-clock uptimes between 68% and 99%, and yielded a frequency value of 444 779 044 095 485.373(44) Hz. With a fractional uncertainty of 9.8× 10 − 17 , it is, to our knowledge, the most accurate optical frequency measurement reported to date. Both frequency values are in agreement with other recent measurements, providing further evidence that the 2021 CIPM recommended frequency value is too high by 1.6 times its uncertainty.
Funding Information
  • European Metrology Programme for Innovation and Research
  • European Union’s Horizon 2020 Research and Innovation Programme
  • European Partnership on Metrology
  • European Union’s Horizon Europe research and innovation programme
  • Research Council of Finland