Optimization of pseudo‐continuous arterial spin labeling at 7T with parallel transmission B1 shimming
Open Access
- 24 August 2021
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 87 (1) , 249-262
Abstract
Purpose: To optimize pseudo‐continuous arterial spin labeling (pCASL) for 7 T, and to further improve the labeling efficiency with parallel RF transmission transmit B1 () shimming.Methods: pCASL parameters were optimized based on field distributions at 7 T with simulation. To increase labeling efficiency, the amplitude at inflowing arteries was increased with parallel RF transmission shimming. The “indv‐shim” with shimming weights calculated for each individual subject, and the “univ‐shim” with universal weights calculated on a group of 12 subjects, were compared with circular polarized (CP) shim. The optimized pCASL sequences with three shimming modes (indv‐shim, univ‐shim, and CP‐shim) were evaluated in 6 subjects who underwent two repeated scans 24 hours apart, along with a pulsed ASL sequence. Quantitative metrics including mean amplitude, perfusion, and intraclass correlation coefficient were calculated. The optimized 7T pCASL was compared with standard 3T pCASL on 5 subjects, using spatial SNR and temporal SNR.Results: The optimal pCASL parameter set (RF duration/gap = 300/250 us, ) achieved robust perfusion measurement in the presence of inhomogeneities. Both indv‐shim and univ‐shim significantly increased amplitude compared with CP‐shim in simulation and in vivo experiment (P < .01). Compared with CP‐shim, perfusion signal was increased by 9.5% with indv‐shim (P < .05) and by 5.3% with univ‐shim (P = .35). All three pCASL sequences achieved fair to good repeatability (intraclass correlation coefficient ≥ 0.5). Compared with 3T pCASL, the optimized 7T pCASL achieved 78.3% higher spatial SNR and 200% higher temporal SNR.Conclusion: The optimized pCASL achieved robust perfusion imaging at 7 T, while both indv‐shim and univ‐shim further increased labeling efficiency.Keywords
This publication has 31 references indexed in Scilit:
- Optimization of adiabatic pulses for pulsed arterial spin labeling at 7 tesla: Comparison with pseudo‐continuous arterial spin labelingMagnetic Resonance in Medicine, 2021
- Pulsed arterial spin labelling at ultra-high field with a B 1 + -optimised adiabatic labelling pulseMagnetic Resonance Materials in Physics, Biology and Medicine, 2016
- Quantitative and functional pulsed arterial spin labeling in the human brain at 9.4 tMagnetic Resonance in Medicine, 2015
- Signal‐to‐noise ratio and MR tissue parameters in human brain imaging at 3, 7, and 9.4 tesla using current receive coil arraysMagnetic Resonance in Medicine, 2015
- In vivo blood T1 measurements at 1.5 T, 3 T, and 7 TMagnetic Resonance in Medicine, 2012
- Feasibility of pseudocontinuous arterial spin labeling at 7 T with whole-brain coverageMagnetic Resonance Materials in Physics, Biology and Medicine, 2011
- Continuous flow‐driven inversion for arterial spin labeling using pulsed radio frequency and gradient fieldsMagnetic Resonance in Medicine, 2008
- A theoretical and experimental investigation of the tagging efficiency of pseudocontinuous arterial spin labelingMagnetic Resonance in Medicine, 2007
- Frequency offset corrected inversion (FOCI) pulses for use in localized spectroscopyMagnetic Resonance in Medicine, 1996
- Adiabatic Pulses for Wideband Inversion and Broadband DecouplingJournal of Magnetic Resonance, Series A, 1995