Pressure‐flow characteristics of breathing systems and their components for pediatric and adult patients
- 17 November 2017
- journal article
- research article
- Published by Wiley in Pediatric Anesthesia
- Vol. 28 (1) , 37-45
Abstract
Summary: Background: Breathing circuits connect the ventilator to the patients’ respiratory system. Breathing tubes, connectors, and sensors contribute to artificial airway resistance to a varying extent. We hypothesized that the flow‐dependent resistance is higher in pediatric breathing systems and their components compared to respective types for adults.Aims: We aimed to characterize the resistance of representative breathing systems and their components used in pediatric patients (including devices for adults) by their nonlinear pressure‐flow relationship.Methods: We used a physical model to measure the flow‐dependent pressure gradient (∆P) across breathing tubes, breathing tube extensions, 90°‐ and Y‐connectors, flow‐ and carbon dioxide sensors, water traps and reusable, disposable and coaxial breathing systems for pediatric and for adult patients. ∆P was analyzed for usual flow ranges and statistically compared at a representative flow rate of 300 mL∙s−1 (∆P300).Results: ∆P across pediatric devices always exceeded ∆P across the corresponding devices for adult patients (all P < .001 [no 95% CI includes 0]). ∆P300 across breathing system components for adults was always below 0.2 cmH2O but reached up to 4.6 cmH2O in a flow sensor for pediatric patients. ∆P300 was considerably higher across the reusable compared to the disposable pediatric breathing systems (1.9 vs 0.3 cmH2O, P < .001, [95% CI −1.59 to −1.56]).Conclusion: The resistances of pediatric breathing systems and their components result in pressure gradients exceeding those for adults several fold. Considering the resistance of individual components is crucial for composing a breathing system matching the patient's needs. Compensation of the additional resistance should be considered if a large composed resistance is unavoidable.Keywords
Funding Information
- Institutional Funding
This publication has 16 references indexed in Scilit:
- Weaning and extubation readiness in pediatric patients*Pediatric Critical Care Medicine, 2009
- Breathing system filters for paediatric practiceAnaesthesia, 2007
- Accuracy of automatic tube compensation in new-generation mechanical ventilators*Critical Care Medicine, 2003
- Continuous calculation of intratracheal pressure in the presence of pediatric endotracheal tubesCritical Care Medicine, 2000
- Intrinsic positive end-expiratory pressure (PEEPi)Intensive Care Medicine, 1995
- Continuous Calculation of Intratracheal Pressure in Tracheally Intubated PatientsAnesthesiology, 1993
- Inspiratory Pressure Support Compensates for the Additional Work of Breathing Caused by the Endotracheal TubeAnesthesiology, 1991
- Assessment of a hygroscopic heat and moisture exchanger for paediatric useAnaesthesia, 1991
- Analysis of Resistance to Gas Flow in Nine Adult Ventilator CircuitsChest, 1989
- Inappropriate triggeringAnaesthesia, 1988