Development of simple high-precision two-dimensional dose-distribution measurement method for proton beam therapy using imaging plate and EBT3
- 28 July 2016
- journal article
- research article
- Published by Springer Nature in Australasian Physics & Engineering Sciences in Medicine
- Vol. 39 (3) , 687-696
Abstract
Although there are several two-dimensional (2D) dose-distribution measurement methods using proton beam therapy, they all have drawbacks; hence, there is no standard method established worldwide. The purpose of this study was to develop a simple, high-precision 2D distribution measurement method for proton beam therapy that uses an imaging plate and EBT3. First, we expanded the maximum readable dose (saturation dose) in the imaging plate. The method involves (i) the control of the fading phenomenon by an annealing process and (ii) the control of the photostimulated luminescence (PSL) phenomenon using a longpass filter (LPF). In method (i), upon heating at 80 °C, the PSL became 0.485 times the room temperature, and in method (ii), we attenuated the PSL by a factor of 0.245 using an LPF. Thus, by combining methods (i) and (ii), we expanded the saturation dose to 2 Gy. Thus, it was possible to measure the imaging plate and EBT3 in the same dose range. We simultaneously measured the percent depth dose using imaging plate and EBT3. We defined a correction factor to match the measured values—which had a reduced sensitivity because of the linear energy transfer (LET) dependence of the imaging plate and EBT3—with reference data and developed a correction factor function. Subsequently, by defining the relative LET dependence of imaging plate and EBT3 as the relative sensitivity and converting the relationship imaging plate between the relative sensitivity and correction factor into a function, we obtained a sensitivity-correction function. By employing this function, measurements with the same accuracy as the reference data were performed using the imaging plate and EBT3.Keywords
This publication has 14 references indexed in Scilit:
- Radiochromic EBT2 film dosimetry for low‐energy protontherapyMedical Physics, 2011
- Dose verification of proton beam therapy using the Gafchromic EBT filmRadiation Measurements, 2011
- Gafchromic® EBT films for ion dosimetryRadiation Measurements, 2010
- Energy dependence and dose response of Gafchromic EBT2 film over a wide range of photon, electron, and proton beam energiesMedical Physics, 2010
- Gafchromic EBT film dosimetry in proton beamsPhysics in Medicine & Biology, 2010
- The comparison of doses measured by radiochromic films and semiconductor detector in a 175MeV proton beamPhysica Medica, 2009
- Accelerator beam data commissioning equipment and procedures: Report of the TG‐106 of the Therapy Physics Committee of the AAPMMedical Physics, 2008
- Use of a two‐dimensional ionization chamber array for proton therapy beam quality assuranceMedical Physics, 2008
- Measurement of depth-dose distribution of protons by an imaging plateNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2003
- Response characteristics of an imaging plate to clinical proton beamsNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 1999