Comparison among TonoVet, TonoVet Plus, Tono-Pen Avia Vet, and Kowa HA-2 portable tonometers for measuring intraocular pressure in dogs
Authors:
- João Victor Goulart Consoni PassareliJPJoão Victor Goulart Consoni Passareli
- Felipe Franco NascimentoFNFelipe Franco Nascimento
- Giovana José Garcia EstanhoGEGiovana José Garcia Estanho
- Claudia Lizandra RicciCRClaudia Lizandra Ricci
- Glaucia Prada KanashiroGKGlaucia Prada Kanashiro
- Rogério GiuffridaRGRogério Giuffrida
- Silvia Franco AndradeSASilvia Franco Andrade
Open Access
- 21 September 2021
- journal article
- research article
- Published by Veterinary World in Veterinary World
- Vol. 14 (9) , 2444-2451
Abstract
Background and Aim: Tonometers are an important instrument for measuring intraocular pressure (IOP) in the diagnosis of glaucoma or uveitis. This study aimed to compare the accuracy of the main types of tonometers with different IOP measurement methodologies in dogs: TonoVet and TonoVet Plus (rebound), Tono-Pen Avia Vet (applanation), and Kowa HA-2 (Goldmann applanation). Materials and Methods: IOP was measured in 152 eyes of 76 dogs. A postmortem study was performed by comparing manometry and tonometry values and calculating the correlation coefficient (r2), in vivo real IOP (manometry) among the tonometers was compared, and an outpatient study was conducted with healthy eyes and eyes with signs of glaucoma and uveitis. Results: In the postmortem study, the values of r2 in descending order were Kowa (0.989), TonoVet Plus (0.984), TonoVet (0.981), and Tono-Pen Avia Vet (0.847). The IOP values in mmHg in the in vivo study were as follows: Aneroid manometer (16.8±2.5.7), TonoVet (18.1±2.9), TonoVet Plus (20.6±2.3), Tono-Pen Avia Vet (17.1±2.5), and Kowa (16.1±1.7); in outpatient clinics: TonoVet (16.8±3.8), TonoVet Plus (19.2±2.9), Tono-Pen Avia Vet (16.2±2.4), and Kowa (15.0±1.3); glaucoma: TonoVet (30.2±3.5), TonoVet Plus (35.0±6.1), Tono-Pen Avia Vet (29.5±4.2), and Kowa (23.9±5.0); and uveitis: TonoVet (14.2±1.4), TonoVet Plus (17.6±1.9), Tono-Pen Avia Vet (13.7±2.1), and Kowa (12.6±1.7). Conclusion: There was a strong correlation between IOP values and manometry in all the tonometers. The highest values were obtained with TonoVet Plus and the lowest with Kowa HA-2. All tonometers accurately measured IOP in dogs, including the latest TonoVet Plus, which showed an excellent correlation coefficient.Keywords
Funding Information
- Coordenação de Aperfeiçoamento de Pessoal de Nível Superior (Financing Code 001)
This publication has 29 references indexed in Scilit:
- A Comparative Study of Rebound Tonometry With Tonopen and Goldmann Applanation Tonometry Following Vitreoretinal SurgeryAmerican Journal of Ophthalmology, 2016
- GlaucomasTopics in Companion Animal Medicine, 2015
- Measuring intraocular pressureCurrent Opinion in Ophthalmology, 2015
- Clinical comparison of theTonoVet®rebound tonometer and theTono‐PenVet®applanation tonometer in dogs and cats with ocular disease: glaucoma or corneal pathologyVeterinary Ophthalmology, 2013
- Comparison of intraocular pressure measurements between the Tono‐Pen XL®and Perkins®applanation tonometers in dogs and catsVeterinary Ophthalmology, 2011
- Evaluation of the Perkins®handheld applanation tonometer in the measurement of intraocular pressure in dogs and catsVeterinary Ophthalmology, 2009
- Comparison of the rebound tonometer (TonoVet®) with the applanation tonometer (TonoPen XL®) in normal Eurasian Eagle owls (Bubo bubo)Veterinary Ophthalmology, 2007
- Comparison of the rebound tonometer (ICare®) to the applanation tonometer (Tonopen XL®) in normotensive dogsVeterinary Ophthalmology, 2006
- Steroid-Induced Ocular Hypertension in Normal CattleArchives of Ophthalmology (1950), 2004
- Tono‐Pen XL®calibration curves for cats, cows and sheepVeterinary Ophthalmology, 2004