Evaluation of the RF depositions at 3T in routine clinical scans with respect to the SAR safety to improve efficiency of MRI utilization
Open Access
- 18 September 2024
- journal article
- research article
- Published by Walter de Gruyter GmbH in Biomedical Engineering / Biomedizinische Technik
- Vol. 70 (1) , 49-59
Abstract
Objectives: This study explores the potential for improving of 3T MRI utilization by assessing and tailoring RF exposure in routine clinical scans while complying to standard safety limit. Methods: Using two generic human body models, we evaluated pbSAR10g values at four landmark positions (knee, pelvis, thoracic spine, head) at different wbSAR levels. Specifically, we analyzed local SAR10g in different operating modes and computed the maximum safety wbSAR, ensuring compliance with IEC limits. Results: In normal operating mode, the RF power deposition reached wbSAR limit before the pbSAR10g limit. In the first level controlled operating mode, pbSAR10g limit is reached before the wbSAR limit in the knee, thoracic spine scanning scenarios, while the wbSAR limit is reached first in the pelvis scanning scenarios, making it the most potential-releasing (up to 33.33 %) scanning scenario. For head exposure, the head SAR10g limit is reached before the wbSAR limit, highlighting the necessity for strict SAR control. Moreover, we calculated the minimum allowable TR for common imaging sequences for reference. Conclusions: Different RF exposure setups are necessary to meet safety standards in various scenarios. By implementing careful RF exposure setups in routine clinical scans, the high potential capacity of 3T MRI can be fully released.Keywords
Funding Information
- Science and Technology Program of Guangdong (Foreign Famous Expert), China (207223078007)
- National Foreign Expert Project (G2023163005L)
- Guangdong Provincial Key Laboratory of Human Digital Twin (2022B1212010004)
- National Natural Science Foundation of China (61671229, 61929101)
This publication has 28 references indexed in Scilit:
- MRI scanner‐independent specific absorption rate measurements using diffusion coefficientsJournal of Applied Clinical Medical Physics, 2017
- An MRI scanner‐independent radiofrequency dosimeter for the estimation of RF power deposition with a human torso phantomConcepts in Magnetic Resonance Part B, 2017
- Specific absorption rate intersubject variability in 7T parallel transmit MRI of the headMagnetic Resonance in Medicine, 2012
- Local specific absorption rate in high‐pass birdcage and transverse electromagnetic body coils for multiple human body models in clinical landmark positions at 3TJournal of Magnetic Resonance Imaging, 2011
- Improved homogeneity of the transmit field by simultaneous transmission with phased array and volume coilJournal of Magnetic Resonance Imaging, 2010
- A prototype RF dosimeter for independent measurement of the average specific absorption rate (SAR) during MRIJournal of Magnetic Resonance Imaging, 2007
- SAR and temperature: Simulations and comparison to regulatory limits for MRIJournal of Magnetic Resonance Imaging, 2007
- Calculations of B1 distribution, SNR, and SAR for a surface coil adjacent to an anatomically‐accurate human body modelMagnetic Resonance in Medicine, 2001
- SAR and B1 field distributions in a heterogeneous human head model within a birdcage coilMagnetic Resonance in Medicine, 1998
- Estimating radiofrequency power deposition in body NMR imagingMagnetic Resonance in Medicine, 1985