Establishment of physiologically relevant oxygen gradients in microfluidic organ chips
Preprint
Abstract
In vitro models of human organs must accurately reconstitute oxygen concentrations and gradients that are observed in vivo to mimic gene expression, metabolism, and host-microbiome interactions. Here we describe a simple strategy to achieve physiologically relevant oxygen tension in a two-channel human small intestine-on-a-chip (Intestine Chip) lined with primary human duodenal epithelium and intestinal microvascular endothelium in parallel channels separated by a porous membrane while both channels are perfused with oxygenated medium. This strategy was developed using computer simulations that predicted lowering the oxygen permeability of poly-dimethlysiloxane (PDMS) chips in specified locations using a gas impermeable film will allow the cells to naturally decrease the oxygen concentration through aerobic respiration and reach steady-state oxygen levels < 36 mm Hg (< 5%) within the epithelial lumen. The approach was experimentally confirmed using chips with embedded oxygen sensors that maintained this stable oxygen gradient. Furthermore, Intestine Chips cultured with this approach supported formation of a villus epithelium interfaced with a continuous endothelium and maintained intestinal barrier integrity for 72 h. This strategy recapitulates in vivo functionality in an efficient, inexpensive, and scalable format that improves the robustness and translatability of Organ Chip technology for studies on microbiome as well as oxygen sensitivity.Keywords
All Related Versions
- Published version: Lab on a Chip, 22 (8), 1584.
This publication has 31 references indexed in Scilit:
- Design Criteria for Generating Physiologically Relevant In Vitro Models in BioreactorsProcesses, 2014
- Optimal periodic perfusion strategy for robust long-term microfluidic cell cultureLab on a Chip, 2013
- Dysbiosis in inflammatory bowel diseases: the oxygen hypothesisThe ISME Journal, 2013
- Mathematical analysis of oxygen transfer through polydimethylsiloxane membrane between double layers of cell culture channel and gas chamber in microfluidic oxygenatorMicrofluidics and Nanofluidics, 2013
- Human gut-on-a-chip inhabited by microbial flora that experiences intestinal peristalsis-like motions and flowLab on a Chip, 2012
- Polyvinyls and Acrylics2012
- FEM-based oxygen consumption and cell viability models for avascular pancreatic isletsTheoretical Biology and Medical Modelling, 2009
- Quantitative measurement and control of oxygen levels in microfluidic poly(dimethylsiloxane) bioreactors during cell cultureBiomedical Microdevices, 2006
- Formation of steady‐state oxygen gradients in vitro: Application to liver zonationBiotechnology and Bioengineering, 2003
- Hyperoxia (145 mmHg pO2) and tissue normoxia (20–40 mmHg pO2) modulate human vascular cell functions1992