Systematic investigation on the validity of partition model dosimetry for 90 Y radioembolization using Monte Carlo simulation
- 22 August 2017
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 62 (18) , 7342-7356
Abstract
We aimed to investigate the validity of the partition model (PM) in estimating the absorbed doses to liver tumour ( D T ), normal liver tissue ( D N L ) and lungs ( D L ), when cross-fire irradiations between these compartments are being considered. MIRD-5 phantom incorporated with various treatment parameters, i.e. tumour involvement (TI), tumour-to-normal liver uptake ratio (T/N) and lung shunting (LS), were simulated using the Geant4 Monte Carlo (MC) toolkit. 10 8 track histories were generated for each combination of the three parameters to obtain the absorbed dose per activity uptake in each compartment ( D T A T , D N L A N L , and D L A L ). The administered activities, A were estimated using PM, so as to achieve either limiting doses to normal liver, D N L l i m or lungs, D L l i m (70 or 30 Gy, respectively). Using these administered activities, the activity uptake in each compartment ( A T , A N L , and A L ) was estimated and multiplied with the absorbed dose per activity uptake attained using the MC simulations, to obtain the actual dose received by each compartment. PM overestimated D L by 11.7% in all cases, due to the escaped particles from the lungs. D T and D N L by MC were largely affected by T/N, which were not considered by PM due to cross-fire exclusion at the tumour-normal liver boundary. These have resulted in the overestimation of D T by up to 8% and underestimation of D N L by as high as −78%, by PM. When D N L l i m was estimated via PM, the MC simulations showed significantly higher D N L for cases with higher T/N, and LS ⩽ 10%. All D L and D T by MC were overestimated by PM, thus D L l i m were never exceeded. PM leads to inaccurate dose estimations due to the exclusion of cross-fire irradiation, i.e. between the tumour and normal liver tissue. Caution should be taken for cases with higher TI and T/N, and lower LS, as they contribute to major underestimation of D N L . For D L , a different correction factor for dose calculation may be used for improved accuracy.Keywords
Funding Information
- Kementerian Sains, Teknologi dan Inovasi (SF011-2014)
- Universiti Malaya (PG104-2013B)
This publication has 31 references indexed in Scilit:
- Recent developments in Geant4Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2016
- Clinical impact of 99mTc-MAA SPECT/CT-based dosimetry in the radioembolization of liver malignancies with 90Y-loaded microspheresEuropean Journal of Nuclear Medicine and Molecular Imaging, 2015
- Personalized Dosimetry with Intensification Using 90Y-Loaded Glass Microsphere Radioembolization Induces Prolonged Overall Survival in Hepatocellular Carcinoma Patients with Portal Vein ThrombosisJournal of Nuclear Medicine, 2015
- Cancer incidence and mortality worldwide: Sources, methods and major patterns in GLOBOCAN 2012International Journal of Cancer, 2014
- Recommendations of the American Association of Physicists in Medicine on dosimetry, imaging, and quality assurance procedures for 90Y microsphere brachytherapy in the treatment of hepatic malignanciesMedical Physics, 2011
- Multimodality imaging can predict the metabolic response of unresectable colorectal liver metastases to radioembolization therapy with Yttrium-90 labeled resin microspheresPhysics in Medicine & Biology, 2008
- Geant4 developments and applicationsIEEE Transactions on Nuclear Science, 2006
- Geant4 low energy electromagnetic physics2005
- Geant4—a simulation toolkitNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2003
- Dose distribution following selective internal radiation therapyInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 1991