Commissioning an Exradin W2 plastic scintillation detector for clinical use in small radiation fields
Open Access
- 21 July 2022
- journal article
- research article
- Published by Wiley in Journal of Applied Clinical Medical Physics
- Vol. 23 (8) , e13728
Abstract
Purpose: The purpose of this work is to evaluate the Standard Imaging Exradin W2 plastic scintillation detector (W2) for use in the types of fields used for stereotactic radiosurgery.Methods: Prior to testing the W2 in small fields, the W2 was evaluated in standard large field conditions to ensure good detector performance. These tests included energy dependence, short‐term repeatability, dose‐response linearity, angular dependence, temperature dependence, and dose rate dependence. Next, scan settings and calibration of the W2 were optimized to ensure high quality data acquisition. Profiles of small fields shaped by cones and multi‐leaf collimator (MLCs) were measured using the W2 and IBA RAZOR diode in a scanning water tank. Output factors for cones (4–17.5 mm) and MLC fields (1, 2, 3 cm) were acquired with both detectors. Finally, the dose at isocenter for seven radiosurgery plans was measured with the W2 detector.Results: W2 exhibited acceptable warm‐up behavior, short‐term reproducibility, axial angular dependence, dose‐rate linearity, and dose linearity. The detector exhibits a dependence upon energy, polar angle, and temperature. Scanning measurements taken with the W2 and RAZOR were in good agreement, with full‐width half‐maximum and penumbra widths agreeing to within 0.1 mm. The output factors measured by the W2 and RAZOR exhibited a maximum difference of 1.8%. For the seven point‐dose measurements of radiosurgery plans, the W2 agreed well with our treatment planning system with a maximum deviation of 2.2%. The Čerenkov light ratio calibration method did not significantly impact the measurement of relative profiles, output factors, or point dose measurements.Conclusion: The W2 demonstrated dosimetric characteristics that are suitable for radiosurgery field measurements. The detector agreed well with the RAZOR diode for output factors and scanned profiles and showed good agreement with the treatment planning system in measurements of clinical treatment plans.Keywords
This publication has 19 references indexed in Scilit:
- Application of the Exradin W1 scintillator to determine Ediode 60017 and microDiamond 60019 correction factors for relative dosimetry within small MV and FFF fieldsPhysics in Medicine & Biology, 2015
- Response to “Comment on ‘Characterization of the Exradin W1 scintillator for use in radiotherapy’ ” [Med. Phys. 42, 297–304 (2015)]Medical Physics, 2015
- Characterization of the Exradin W1 scintillator for use in radiotherapyMedical Physics, 2015
- Variation of kQclin,Qmsrfclin,fmsr for the small‐field dosimetric parameters percentage depth dose, tissue‐maximum ratio, and off‐axis ratioMedical Physics, 2014
- Dosimetric characterization of the Exradin W1 plastic scintillator detector through comparison with an in-house developed scintillator systemRadiation Measurements, 2014
- On the correction, perturbation and modification of small field detectors in relative dosimetryPhysics in Medicine & Biology, 2014
- Application of Cerenkov radiation generated in plastic optical fibers for therapeutic photon beam dosimetryJournal of Biomedical Optics, 2013
- Spectral method for the correction of the Cerenkov light effect in plastic scintillation detectors: A comparison study of calibration procedures and validation in Cerenkov light‐dominated situationsMedical Physics, 2011
- Spectral discrimination of Čerenkov radiation in scintillating dosimetersMedical Physics, 2005
- Scintillating fiber dosimeter for radiation therapy acceleratorIEEE Transactions on Nuclear Science, 2002