Quantification of local image noise variation in PET images for standardization of noise-dependent analysis metrics
- 23 February 2017
- journal article
- Published by IOP Publishing in Biomedical Physics & Engineering Express
- Vol. 3 (2) , 025007
Abstract
Purpose. Analysis metrics such as tumor-to-background (T/B) ratios that rely indirectly on local noise properties of a PET image are known to be highly variable depending on image acquisition and image reconstruction parameters. Less well understood are the effects on local image noise arising from correction algorithms that compensate for photon attenuation and detector sensitivity (attenuation and sensitivity corrections). This work investigates these geometry-dependent local noise characteristics in positron emission tomography (PET) images and their implications for the analysis metrics. Methods. Two different phantoms containing homogeneous radioactivity concentrations were scanned for various initial activities and scan times and then reconstructed with a range of reconstruction parameters. Local noise in the form of a local standard deviation was computed and its spatial variation quantified across the PET imaging volume. Results. The local noise variation across the PET image arising from attenuation and sensitivity corrections can be understood with relatively simple analytical models. For fixed reconstruction parameters, the local standard deviation varied up to 60% in the axial direction and up to 30% in trans-axial (radial) direction. Varying the initial activity concentration and the scan time confirmed that the local noise obeys almost perfect Poisson statistics. As a consequence, depending on the spatial location of the volumes of interest, the T/B metrics can vary more than 30%. In addition, the local standard deviation showed substantial variation up to an order of magnitude (factor 10) for different reconstruction parameters. Conclusion. Quantifying the spatial variation of the geometry-dependent local noise across the imaging volume of a particular PET scanner allows us to develop quantitative correction algorithms for standardizing image analysis using noise-dependent metrics.Keywords
This publication has 23 references indexed in Scilit:
- PET-based quantification of statistical properties of hypoxic tumor subvolumes in head and neck cancerPhysica Medica, 2015
- FDG PET/CT: EANM procedure guidelines for tumour imaging: version 2.0European Journal of Nuclear Medicine and Molecular Imaging, 2014
- PET/CT imaging in cancer: Current applications and future directionsCancer, 2014
- Harmonizing SUVs in multicentre trials when using different generation PET systems: prospective validation in non-small cell lung cancer patientsEuropean Journal of Nuclear Medicine and Molecular Imaging, 2013
- Mutatis Mutandis: Harmonize the Standard!Journal of Nuclear Medicine, 2011
- Effects of injected dose, BMI and scanner type on NECR and image noise in PET imagingPhysics in Medicine & Biology, 2011
- SUVref: reducing reconstruction-dependent variation in PET SUVEJNMMI Research, 2011
- FDG PET and PET/CT: EANM procedure guidelines for tumour PET imaging: version 1.0European Journal of Nuclear Medicine and Molecular Imaging, 2009
- High resolution PET, SPECT and projection imaging in small animalsComputerized Medical Imaging and Graphics, 2001
- Noise propagation in SPECT images reconstructed using an iterative maximum-likelihood algorithmPhysics in Medicine & Biology, 1993